Effects of tube voltage, radiation dose and adaptive statistical iterative reconstruction strength level on the detection and characterization of pulmonary nodules in ultra-low-dose chest CT
Detection rates are similar between 70 kV and 100 kV scans. The 70 kV images have better noise performance under the same ASiR-V level, while images of 100 kV and higher ASiR-V levels are better in preserving the nodule morphology (lower DC and SP); the dose levels above 0.33mSv provide high sensitivity for nodules detection, especially the simulated ground glass nodules.