Abstract A008: The effect of body mass index on Tumor Treating Fields (TTFields) intensity distribution in the abdomen: results of a simulation model study

Abstract Introduction: Tumor Treating Fields (TTFields) are electric fields that disrupt processes critical for cancer cell viability and tumor progression. They are generated by a portable medical device and delivered noninvasively to the tumor site via skin-placed arrays. TTFields therapy is approved for the treatment of recurrent and newly diagnosed glioblastoma as well as pleural mesothelioma. Recently, the randomized, pivotal (phase 3) LUNAR study (NCT02973789) demonstrated improved overall survival for TTFields therapy (150 kHz) delivered continuously until progression or intolerable toxicity together with an immune checkpoint inhibitor (ICI) or docetaxel (DTX) vs ICI/DTX alone in patients with metastatic non-small cell lung cancer following progression on or after platinum-based therapy. PANOVA-3 (NCT03377491) is an ongoing pivotal (phase 3), randomized, open-label study evaluating the safety and efficacy of TTFields therapy (150 kHz) together with nab-paclitaxel and gemcitabine for the treatment of patients with unresectable, locally advanced pancreatic adenocarcinoma. The effects of TTFields are dependent on several factors (i.e., intensity, frequency, and distribution) which may be affected by tissue composition/conductivity and array layout. Here, we evaluate guideline TTFields array layouts for TTFields delivery to the abdominal region at therapeutic intensities (∼1 V/cm [amplitude]) using simulation models with varying body mass index (BMI). Methods: Computerized phantom models (Ella, females with BMIs of 22, 26 and 30 kg/m2) were used to simulate the distribution of TTFields. Two different arrays sizes (small [13 disks] and large [20 disks]) were applied to the abdominal region of the models based on abdominal TTFields array layout guidelines. Region of interest (ROIs) included the left and right hypochondriac, left and right lumbar, epigastric, and umbilical regions. Simulations were performed using the Sim4Life platform v6.2.1.4910 (Zurich MedTech, Zurich, Switzerland). Electric field intensity distribution maps were generated for each model, and electric field intensity values across all ROIs in the abdomen were analyzed. Results: Electric field intensities were generally lower in the higher BMI model compared with the lower BMI models (field intensities across all ROIs were 1.60–2.57 V/cm for the 22 kg/m2 BMI model, 1.36–2.13 V/cm in the 26 kg/m2 BMI model and 1.35–1.97 V/cm in the 30 kg/m2 BMI model). However, all BMI models achieved TTFields intensities at therapeutic levels (∼1 V/cm) in all ROIs. Conclusions: Results from this simulation-based study demonstrate the feasibility of delivering TTFields at therapeutic intensities to patients with abdominal tumors regardless of their BMI when using abdominal guideline layouts. Citation Format: Ariel Naveh, Nadav Shapira. The effect of body mass index on Tumor Treating Fields (TTFields) intensity distribution in the abdomen: results of a simulation model study [abstract]. In: Proceedings of the AACR Special Conference in Cancer Research: Advances in Pancreatic Cancer Research; 2024 Sep 15-18; Boston, MA. Philadelphia (PA): AACR; Cancer Res 2024;84(17 Suppl_2):Abstract nr A008.

Authors

Publication Details

Journal
Cancer Research
Published
2024-09-15
DOI
https://doi.org/10.1158/1538-7445.pancreatic24-a008
Primary Topic
Radiomics and Machine Learning in Medical Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Abstract A008: The effect of body mass index on Tumor Treating Fields (TTFields) intensity distribution in the abdomen: results of a simulation model study

Nadav Shapira, Ariel Naveh
Cancer Research
Radiomics and Machine Learning in Medical Imaging
article

Abstract A008: The effect of body mass index on Tumor Treating Fields (TTFields) intensity distribution in the abdomen: results of a simulation model study

Nadav Shapira, Ariel Naveh
article en

Abstract

Abstract Introduction: Tumor Treating Fields (TTFields) are electric fields that disrupt processes critical for cancer cell viability and tumor progression. They are generated by a portable medical device and delivered noninvasively to the tumor site via skin-placed arrays. TTFields therapy is approved for the treatment of recurrent and newly diagnosed glioblastoma as well as pleural mesothelioma. Recently, the randomized, pivotal (phase 3) LUNAR study (NCT02973789) demonstrated improved overall survival for TTFields therapy (150 kHz) delivered continuously until progression or intolerable toxicity together with an immune checkpoint inhibitor (ICI) or docetaxel (DTX) vs ICI/DTX alone in patients with metastatic non-small cell lung cancer following progression on or after platinum-based therapy. PANOVA-3 (NCT03377491) is an ongoing pivotal (phase 3), randomized, open-label study evaluating the safety and efficacy of TTFields therapy (150 kHz) together with nab-paclitaxel and gemcitabine for the treatment of patients with unresectable, locally advanced pancreatic adenocarcinoma. The effects of TTFields are dependent on several factors (i.e., intensity, frequency, and distribution) which may be affected by tissue composition/conductivity and array layout. Here, we evaluate guideline TTFields array layouts for TTFields delivery to the abdominal region at therapeutic intensities (∼1 V/cm [amplitude]) using simulation models with varying body mass index (BMI). Methods: Computerized phantom models (Ella, females with BMIs of 22, 26 and 30 kg/m2) were used to simulate the distribution of TTFields. Two different arrays sizes (small [13 disks] and large [20 disks]) were applied to the abdominal region of the models based on abdominal TTFields array layout guidelines. Region of interest (ROIs) included the left and right hypochondriac, left and right lumbar, epigastric, and umbilical regions. Simulations were performed using the Sim4Life platform v6.2.1.4910 (Zurich MedTech, Zurich, Switzerland). Electric field intensity distribution maps were generated for each model, and electric field intensity values across all ROIs in the abdomen were analyzed. Results: Electric field intensities were generally lower in the higher BMI model compared with the lower BMI models (field intensities across all ROIs were 1.60–2.57 V/cm for the 22 kg/m2 BMI model, 1.36–2.13 V/cm in the 26 kg/m2 BMI model and 1.35–1.97 V/cm in the 30 kg/m2 BMI model). However, all BMI models achieved TTFields intensities at therapeutic levels (∼1 V/cm) in all ROIs. Conclusions: Results from this simulation-based study demonstrate the feasibility of delivering TTFields at therapeutic intensities to patients with abdominal tumors regardless of their BMI when using abdominal guideline layouts. Citation Format: Ariel Naveh, Nadav Shapira. The effect of body mass index on Tumor Treating Fields (TTFields) intensity distribution in the abdomen: results of a simulation model study [abstract]. In: Proceedings of the AACR Special Conference in Cancer Research: Advances in Pancreatic Cancer Research; 2024 Sep 15-18; Boston, MA. Philadelphia (PA): AACR; Cancer Res 2024;84(17 Suppl_2):Abstract nr A008.

Cancer ResearchVol. 84(17_Supplement_2)
Openalex Percentile: Top 74%
Radiomics and Machine Learning in Medical Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.