Measuring total hip arthroplasty liner wear using the EOS Imaging System: experimental and clinical results
Experimental THA liner wear using EOS was within clinically relevant tolerances and without bias. The clinical precision was just outside our defined clinically relevant precision. Compared with conventional RSA, EOS is less accurate and precise but may still be of value for certain clinical applications, provided larger sample size or longer follow-up are available.