In-Vivo Dosimetry for Ultra-High Dose Rate (UHDR) Electron Beam FLASH Radiotherapy Using an Organic (Plastic), an Organic–Inorganic Hybrid and an Inorganic Point Scintillator System

Dosimetry is crucial in radiotherapy to warrant safe and correct treatment. In FLASH radiotherapy, where ultra-high dose rates (UHDRs) are used, the dosimetric demands are more stringent, requiring the development and investigation of new dosemeters. In this study, three prototype fiber-optic dosemeters (FODs)—an inorganic, an organic–inorganic hybrid metal halide, and an organic (plastic) scintillator are optimized and investigated for UHDR electron irradiations. The plastic FOD is developed by Medscint, whereas the others are in-house made. The stem signal is minimized by spectral decomposition for the plastic scintillator, and by band-pass wavelength filters for the inorganic and organic–inorganic hybrid metal halide FOD. All prototypes are tested for the dose rate defining parameters. The optimal band-pass wavelength filters are found to be centered around 500 nm and 425 nm for the inorganic and organic–inorganic hybrid metal halide FODs, respectively. A sampling frequency of 1000 Hz is chosen for the inorganic and organic–inorganic hybrid metal halide FODs. The plastic FOD shows to be the least dose rate dependent with maximum deviations of 3% from the reference for the relevant beam settings. The inorganic and organic–inorganic hybrid metal halide FODs, in contrast, show large deviations of >10% from the reference and require more investigation. The current FOD prototypes are insufficient for application in UHDR electron beams, and require further development and investigation.

Authors

Institutions

Publication Details

Journal
Photonics
Published
2024-09-13
DOI
https://doi.org/10.3390/photonics11090865
Citations
5
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
3.04
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

In-Vivo Dosimetry for Ultra-High Dose Rate (UHDR) Electron Beam FLASH Radiotherapy Using an Organic (Plastic), an Organic–Inorganic Hybrid and an Inorganic Point Scintillator System

Paul Sterckx, Paul Leblans, Verdi Vanreusel, Luana de Freitas Nascimento et al.
5 citations
Photonics
Advanced Radiotherapy Techniques
3.04
article

In-Vivo Dosimetry for Ultra-High Dose Rate (UHDR) Electron Beam FLASH Radiotherapy Using an Organic (Plastic), an Organic–Inorganic Hybrid and an Inorganic Point Scintillator System

Paul Sterckx, Paul Leblans, Verdi Vanreusel, Luana de Freitas Nascimento, B. M. Cote, Dirk Verellen, Hugo Vallet, Dirk Vandenbroucke, J. Wijnen
article en
5 citations

Abstract

Dosimetry is crucial in radiotherapy to warrant safe and correct treatment. In FLASH radiotherapy, where ultra-high dose rates (UHDRs) are used, the dosimetric demands are more stringent, requiring the development and investigation of new dosemeters. In this study, three prototype fiber-optic dosemeters (FODs)—an inorganic, an organic–inorganic hybrid metal halide, and an organic (plastic) scintillator are optimized and investigated for UHDR electron irradiations. The plastic FOD is developed by Medscint, whereas the others are in-house made. The stem signal is minimized by spectral decomposition for the plastic scintillator, and by band-pass wavelength filters for the inorganic and organic–inorganic hybrid metal halide FOD. All prototypes are tested for the dose rate defining parameters. The optimal band-pass wavelength filters are found to be centered around 500 nm and 425 nm for the inorganic and organic–inorganic hybrid metal halide FODs, respectively. A sampling frequency of 1000 Hz is chosen for the inorganic and organic–inorganic hybrid metal halide FODs. The plastic FOD shows to be the least dose rate dependent with maximum deviations of 3% from the reference for the relevant beam settings. The inorganic and organic–inorganic hybrid metal halide FODs, in contrast, show large deviations of >10% from the reference and require more investigation. The current FOD prototypes are insufficient for application in UHDR electron beams, and require further development and investigation.

PhotonicsVol. 11(9)
University of Antwerp (BE), Agfa-Gevaert (Belgium) (BE), Québec Metro High Tech Park (Canada) (CA), Belgian Nuclear Research Centre (BE), Iridium Kankernetwerk (BE)
Openalex Percentile: Top 9%
Advanced Radiotherapy Techniques
3.04
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.