A novel mechanical design of a bolometric array for the CROSS double-beta decay experiment

Abstract The CROSS experiment will search for neutrinoless double-beta decay using a specific mechanical structure to hold thermal detectors. The design of the structure was tuned to minimize the background contribution, keeping an optimal detector performance. A single module of the structure holds two scintillating bolometers (with a crystal size of 45 × 45 × 45 mm and a Ge slab facing the crystal's upper side) in the Cu frame, allowing for a modular construction of a large-scale array. Two designs are released: the initial Thick version contains around 15% of Cu over the crystal mass (lithium molybdate, LMO), while this ratio is reduced to ∼ 6% in a finer ( Slim ) design. Both designs were tested extensively at aboveground (IJCLab, France) and underground (LSC, Spain) laboratories. In particular, at LSC we used a pulse-tube-based CROSS facility to operate a 6-crystal array of LMOs enriched/depleted in 100 Mo. The tested LMOs show high spectrometric performance in both designs; notably, the measured energy resolution is 5–7 keV FWHM at 2615 keV γ s, nearby the Q-value of 100 Mo (3034 keV). Due to the absence of a reflective cavity around LMOs, a low scintillation signal is detected by Ge bolometers: ∼ 0.3 keV (150 photons) for 1-MeV γ ( β ) LMO-event. Despite that, an acceptable separation between α and γ ( β events is achieved with most devices. The highest efficiency is reached with light detectors in the Thick design thanks to a lower baseline noise width (0.05–0.09 keV RMS) when compared to that obtained in the Slim version (0.10–0.35 keV RMS). Given the pivotal role of bolometric photodetectors for particle identification and random coincidences rejection, we will use the structure here described with upgraded light detectors, featuring thermal signal amplification via the Neganov-Trofimov-Luke effect, as also demonstrated in the present work.

Authors

Institutions

Publication Details

Journal
Journal of Instrumentation
Published
2024-09-01
DOI
https://doi.org/10.1088/1748-0221/19/09/p09014
Citations
6
Primary Topic
Neutrino Physics Research
Type
article
Field-Weighted Citation Impact
3.96

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A novel mechanical design of a bolometric array for the CROSS double-beta decay experiment

A. Gallas, D. V. Poda, S.I. Konovalov, L. Imbert et al.
6 citations
Journal of Instrumentation
Neutrino Physics Research
3.96
article

A novel mechanical design of a bolometric array for the CROSS double-beta decay experiment

A. Gallas, D. V. Poda, S.I. Konovalov, L. Imbert, V. Berest, I. Dafinei, P. Loaiza, M. M. Zarytskyy, J. M. Calvo-Mozota, F.A. Danevich, Claire A. Marrache-Kikuchi, Aldo Ianni, V. Kobychev, G. Pessina, P. Gras, Y. Peinaud, S. Marnieros, C. Gotti, V. I. Umatov, V.I. Tretyak, T. Dixon, P. Carniti, A. S. Barabash, A. Giuliani, F. Ferri, E. Olivieri, A. Zolotarova, M. Martínez, Ph. Rosier, H. Khalife, C. Nones, L. Dumoulin, P. de Marcillac, D. Auguste, L. Bergé, J. A. Scarpaci, A. Órtiz de Solórzano, M. Chapellier
article en
6 citations

Abstract

Abstract The CROSS experiment will search for neutrinoless double-beta decay using a specific mechanical structure to hold thermal detectors. The design of the structure was tuned to minimize the background contribution, keeping an optimal detector performance. A single module of the structure holds two scintillating bolometers (with a crystal size of 45 × 45 × 45 mm and a Ge slab facing the crystal's upper side) in the Cu frame, allowing for a modular construction of a large-scale array. Two designs are released: the initial Thick version contains around 15% of Cu over the crystal mass (lithium molybdate, LMO), while this ratio is reduced to ∼ 6% in a finer ( Slim ) design. Both designs were tested extensively at aboveground (IJCLab, France) and underground (LSC, Spain) laboratories. In particular, at LSC we used a pulse-tube-based CROSS facility to operate a 6-crystal array of LMOs enriched/depleted in 100 Mo. The tested LMOs show high spectrometric performance in both designs; notably, the measured energy resolution is 5–7 keV FWHM at 2615 keV γ s, nearby the Q-value of 100 Mo (3034 keV). Due to the absence of a reflective cavity around LMOs, a low scintillation signal is detected by Ge bolometers: ∼ 0.3 keV (150 photons) for 1-MeV γ ( β ) LMO-event. Despite that, an acceptable separation between α and γ ( β events is achieved with most devices. The highest efficiency is reached with light detectors in the Thick design thanks to a lower baseline noise width (0.05–0.09 keV RMS) when compared to that obtained in the Slim version (0.10–0.35 keV RMS). Given the pivotal role of bolometric photodetectors for particle identification and random coincidences rejection, we will use the structure here described with upgraded light detectors, featuring thermal signal amplification via the Neganov-Trofimov-Luke effect, as also demonstrated in the present work.

Journal of InstrumentationVol. 19(09)
Commissariat à l'Énergie Atomique et aux Énergies Alternatives (FR), CEA Paris-Saclay (FR), Laboratoire de Physique des 2 Infinis Irène Joliot-Curie (FR), Institut de Recherche sur les Lois Fondamentales de l'Univers (FR)
National Research Foundation of Ukraine, European Commission, National Research Foundation, Agence Nationale de la Recherche
Affordable and clean energy
Openalex Percentile: Top 5%
Neutrino Physics Research
3.96
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.