Evaluation of cobas® Liat® Cdiff, STANDARD™ M10 C. difficile and Xpert® C. difficile BT assays for rapid detection of toxigenic Clostridioides difficile
We evaluated the analytical performance of three commercial molecular assays for rapid detection of Clostridioides difficile toxin B in stool samples. The results were compared with results from the BD MAX™ Cdiff assay. We analyzed forty negative and thirty-two positive stool samples with three rapid assays: Roche cobas® Liat® Cdiff, SD Biosensor STANDARD™ M10 C. difficile and Cepheid Xpert® C. difficile BT. The assays demonstrated a sensitivity of 96.9 %, 96.9 % and 100.0 % and a specificity of 100 %, 97.5 % and 97.5 %, respectively. There is limited data available on the analytical performance of the newly introduced STANDARD™ M10 C. difficile assay. In this study, all three rapid assays demonstrated similarly high analytical performance and can be used for detection of toxigenic C. difficile.
Authors
- Juulia Suominen
- Suvi Korhonen
- Hanna Jarva (ORCID: https://orcid.org/0000-0002-9154-354X)
- Anu Pätäri‐Sampo (ORCID: https://orcid.org/0000-0001-6714-3688)
- Heini Kutvonen
- Raisa Loginov
Institutions
- University of Helsinki (FI)
- Helsinki University Hospital (FI)
- Hospital District of Helsinki and Uusimaa (FI)
Publication Details
- Journal
- Journal of Microbiological Methods
- Published
- 2024-09-13
- DOI
- https://doi.org/10.1016/j.mimet.2024.107043
- Primary Topic
- Clostridium difficile and Clostridium perfringens research
- Type
- article
- Field-Weighted Citation Impact
- 0.00