A Reduced Resistance, Concentric-Gated Artificial Membrane Lung for Pediatric End-Stage Lung Failure

The goal of the low-resistance pediatric artificial lung (PAL-LR) is to serve as a pumpless bridge-to-transplant device for children with end-stage lung failure. The PAL-LR doubles the exposed fiber length of the previous PAL design. In vitro and in vivo studies tested hemocompatibility, device flow, gas exchange and pressure drop performance. For in vitro tests, average rated blood flow (outlet SO 2 of 95%) was 2.56 ± 0.93 L/min with a pressure drop of 25.88 ± 0.90 mm Hg. At the targeted pediatric flow rate of 1 L/min, the pressure drop was 8.6 mm Hg compared with 25 mm Hg of the PAL. At rated flow, the average O 2 and CO 2 transfer rates were 101.75 ± 10.81 and 77.93 ± 8.40 mL/min, respectively. The average maximum O 2 and CO 2 exchange efficiencies were 215.75 ± 22.93 and 176.99 ± 8.40 mL/(min m 2 ), respectively. In vivo tests revealed an average outlet SO 2 of 100%, and average pressure drop of 2 ± 0 mm Hg for a blood flow of 1.07 ± 0.02 L/min. Having a lower resistance, the PAL-LR is a promising step closer to a pumpless artificial membrane lung that alleviates right ventricular strain associated with idiopathic pulmonary hypertension.

Authors

Institutions

Publication Details

Journal
ASAIO Journal
Published
2024-09-13
DOI
https://doi.org/10.1097/mat.0000000000002308
Primary Topic
Mechanical Circulatory Support Devices
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A Reduced Resistance, Concentric-Gated Artificial Membrane Lung for Pediatric End-Stage Lung Failure

Álvaro Rojas-Peña, Leah M. Gudex, Joseph A. Potkay, Lauren A. Wich et al.
ASAIO Journal
Mechanical Circulatory Support Devices
article

A Reduced Resistance, Concentric-Gated Artificial Membrane Lung for Pediatric End-Stage Lung Failure

Álvaro Rojas-Peña, Leah M. Gudex, Joseph A. Potkay, Lauren A. Wich, Alex J. Thompson, Hannah J. Matich, Tyler Dann, Michael Atie, Matthew Johnson, Robert H. Bartlett, Ronald B. Hirschl
article en

Abstract

The goal of the low-resistance pediatric artificial lung (PAL-LR) is to serve as a pumpless bridge-to-transplant device for children with end-stage lung failure. The PAL-LR doubles the exposed fiber length of the previous PAL design. In vitro and in vivo studies tested hemocompatibility, device flow, gas exchange and pressure drop performance. For in vitro tests, average rated blood flow (outlet SO 2 of 95%) was 2.56 ± 0.93 L/min with a pressure drop of 25.88 ± 0.90 mm Hg. At the targeted pediatric flow rate of 1 L/min, the pressure drop was 8.6 mm Hg compared with 25 mm Hg of the PAL. At rated flow, the average O 2 and CO 2 transfer rates were 101.75 ± 10.81 and 77.93 ± 8.40 mL/min, respectively. The average maximum O 2 and CO 2 exchange efficiencies were 215.75 ± 22.93 and 176.99 ± 8.40 mL/(min m 2 ), respectively. In vivo tests revealed an average outlet SO 2 of 100%, and average pressure drop of 2 ± 0 mm Hg for a blood flow of 1.07 ± 0.02 L/min. Having a lower resistance, the PAL-LR is a promising step closer to a pumpless artificial membrane lung that alleviates right ventricular strain associated with idiopathic pulmonary hypertension.

ASAIO JournalVol. 71(3)
University of Michigan (US)
Good health and well-being
Openalex Percentile: Top 86%
Mechanical Circulatory Support Devices
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.