Effect of mirror system and scanner bed of a flatbed scanner on lateral response artefact in radiochromic film dosimetry

Radiochromic film, evaluated with flatbed scanners, is used for practical radiotherapy QA dosimetry. Film and scanner component effects contribute to the Lateral Response Artefact (LRA), which is further enhanced by light polarisation from both. This study investigates the scanner bed's contribution to LRA and also polarisation from the mirrors for widely used EPSON scanners, as part of broader investigations of this dosimetry method aiming to improve processes and uncertainties. Alternative scanner bed materials were compared on a modified EPSON V700 scanner. Polarisation effects were investigated for complete scanners (V700, V800, on- and off-axis, and V850 on-axis), for a removed V700 mirror system, and independently using retail-quality single mirror combinations simulating practical scanner arrangements, but with varying numbers (0-5) and angles. Some tests had no film present, whilst others included films (EBT3) irradiated to 6 MV doses of 0-11.3 Gy. For polarisation analysis, images were captured by a Canon 7D camera with 50 mm focal length lens. Different scanner bed materials showed only small effects, within a few percent, indicating that the normal glass bed is a good choice. Polarisation varied with scanner type (7-11%), increasing at 10 cm lateral off-axis distance by around a further 6%, and also with film dose. The V700 mirror system showed around 2% difference to the complete scanner. Polarization increased with number of mirrors in the single mirror combinations, to 14% for 4 and 5 mirrors, but specific values depend on angles and mirror quality. Novel film measurement methods could reduce LRA effect corrections and associated uncertainties.

Authors

Institutions

Publication Details

Journal
Physical and Engineering Sciences in Medicine
Published
2024-09-12
DOI
https://doi.org/10.1007/s13246-024-01478-x
Citations
1
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.61

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Effect of mirror system and scanner bed of a flatbed scanner on lateral response artefact in radiochromic film dosimetry

Martin J Butson, Nick Bennie, David Thwaites, Tarafder Shameem
1 citations
Physical and Engineering Sciences in Medicine
Advanced Radiotherapy Techniques
0.61
article

Effect of mirror system and scanner bed of a flatbed scanner on lateral response artefact in radiochromic film dosimetry

Martin J Butson, Nick Bennie, David Thwaites, Tarafder Shameem
article en
1 citations

Abstract

Radiochromic film, evaluated with flatbed scanners, is used for practical radiotherapy QA dosimetry. Film and scanner component effects contribute to the Lateral Response Artefact (LRA), which is further enhanced by light polarisation from both. This study investigates the scanner bed's contribution to LRA and also polarisation from the mirrors for widely used EPSON scanners, as part of broader investigations of this dosimetry method aiming to improve processes and uncertainties. Alternative scanner bed materials were compared on a modified EPSON V700 scanner. Polarisation effects were investigated for complete scanners (V700, V800, on- and off-axis, and V850 on-axis), for a removed V700 mirror system, and independently using retail-quality single mirror combinations simulating practical scanner arrangements, but with varying numbers (0-5) and angles. Some tests had no film present, whilst others included films (EBT3) irradiated to 6 MV doses of 0-11.3 Gy. For polarisation analysis, images were captured by a Canon 7D camera with 50 mm focal length lens. Different scanner bed materials showed only small effects, within a few percent, indicating that the normal glass bed is a good choice. Polarisation varied with scanner type (7-11%), increasing at 10 cm lateral off-axis distance by around a further 6%, and also with film dose. The V700 mirror system showed around 2% difference to the complete scanner. Polarization increased with number of mirrors in the single mirror combinations, to 14% for 4 and 5 mirrors, but specific values depend on angles and mirror quality. Novel film measurement methods could reduce LRA effect corrections and associated uncertainties.

Physical and Engineering Sciences in MedicineVol. 47(4)
The University of Sydney (AU), Taronga Conservation Society Australia (AU), Cancer Institute of New South Wales (AU), Northern Cancer Institute (AU)
University of Sydney
Openalex Percentile: Top 31%
Advanced Radiotherapy Techniques
0.61
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.