First in-Lab Testing of a Cost-Effective Prototype for PM2.5 Monitoring: The P.ALP Assessment

The goal of the present research was to assess, under controlled laboratory conditions, the accuracy and precision of a prototype device (named 'P.ALP': Ph.D. Air-quality Low-cost Project) developed for PM<sub>2.5</sub> concentration level monitoring. Indeed, this study follows a complementary manuscript (previously published) focusing on the in-field evaluation of the device's performance. Four P.ALP prototypes were co-located with the reference instrument in a calm-air aerosol chamber at the NIOSH laboratories in Pittsburgh, PA (USA), used by the Center for Direct Reading and Sensor Technologies. The devices were tested for 10 monitoring days under several exposure conditions. To evaluate the performance of the prototypes, different approaches were employed. After the data from the devices were stored and prepared for analysis, to assess the accuracy (comparing the reference instrument with the prototypes) and the precision (comparing all the possible pairs of devices) of the P.ALPs, linear regression analysis was performed. Moreover, to find out the applicability field of this device, the US EPA's suggested criteria were adopted, and to assess error trends of the prototype in the process of data acquisition, Bland-Altman plots were built. The findings show that, by introducing ad hoc calibration factors, the P.ALP's performance needs to be further implemented, but the device can monitor the concentration trend variations with satisfying accuracy. Overall, the P.ALP can be involved in and adapted to a wide range of applications because of the inexpensive nature of the components, the small dimensions, and the high data storage capacity.

Authors

Institutions

Publication Details

Journal
Sensors
Published
2024-09-12
DOI
https://doi.org/10.3390/s24185915
Citations
1
Primary Topic
Air Quality Monitoring and Forecasting
Type
article
Field-Weighted Citation Impact
0.21

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

First in-Lab Testing of a Cost-Effective Prototype for PM2.5 Monitoring: The P.ALP Assessment

Justin Patts, Andrea Cattaneo, Andrea Spinazzè, Domenico Maria Cavallo et al.
1 citations
Sensors
Air Quality Monitoring and Forecasting
0.21
article

First in-Lab Testing of a Cost-Effective Prototype for PM2.5 Monitoring: The P.ALP Assessment

Justin Patts, Andrea Cattaneo, Andrea Spinazzè, Domenico Maria Cavallo, Emanuele Cauda, Davide Campagnolo, Cody M. Wolfe, Giacomo Fanti, Francesca Borghi
article en
1 citations

Abstract

The goal of the present research was to assess, under controlled laboratory conditions, the accuracy and precision of a prototype device (named 'P.ALP': Ph.D. Air-quality Low-cost Project) developed for PM<sub>2.5</sub> concentration level monitoring. Indeed, this study follows a complementary manuscript (previously published) focusing on the in-field evaluation of the device's performance. Four P.ALP prototypes were co-located with the reference instrument in a calm-air aerosol chamber at the NIOSH laboratories in Pittsburgh, PA (USA), used by the Center for Direct Reading and Sensor Technologies. The devices were tested for 10 monitoring days under several exposure conditions. To evaluate the performance of the prototypes, different approaches were employed. After the data from the devices were stored and prepared for analysis, to assess the accuracy (comparing the reference instrument with the prototypes) and the precision (comparing all the possible pairs of devices) of the P.ALPs, linear regression analysis was performed. Moreover, to find out the applicability field of this device, the US EPA's suggested criteria were adopted, and to assess error trends of the prototype in the process of data acquisition, Bland-Altman plots were built. The findings show that, by introducing ad hoc calibration factors, the P.ALP's performance needs to be further implemented, but the device can monitor the concentration trend variations with satisfying accuracy. Overall, the P.ALP can be involved in and adapted to a wide range of applications because of the inexpensive nature of the components, the small dimensions, and the high data storage capacity.

SensorsVol. 24(18)
University of Insubria (IT), Centers for Disease Control and Prevention (US), National Institute for Occupational Safety and Health (US), University of Bologna (IT)
National Institute for Occupational Safety and Health
Openalex Percentile: Top 51%
Air Quality Monitoring and Forecasting
0.21
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.