Interference from anti-drug antibodies on the quantification of insulin: a comparison of an LC-MS/MS assay and immunoassays

<b>Aim:</b> This study aims to compare the anti-drug antibody (ADA) interference in four pharmacokinetic (PK) assays across different platforms (AlphaLISA, Gyrolab, LC-MS/MS) and to devise a strategy for ADA interference mitigation to improve the accuracy of measured drug in total PK assays.<b>Materials & methods:</b> Spiked test samples, created to achieve different ADA concentrations in human serum also containing an insulin analogue, were analyzed alongside pooled clinical samples using four assays.<b>Results & conclusion:</b> Interference was observed in all platforms. A novel approach using the Gyrolab mixing CD, including acid dissociation in the PK assay, significantly reduced interference and thereby improved relative error from >99% to ≤20% yielding measurements well within the acceptance criteria. Clinical sample results reinforced findings from the test samples.

Authors

Institutions

Publication Details

Journal
Bioanalysis
Published
2024-09-12
DOI
https://doi.org/10.1080/17576180.2024.2389637
Citations
1
Primary Topic
Biosimilars and Bioanalytical Methods
Type
article
Field-Weighted Citation Impact
0.24
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Interference from anti-drug antibodies on the quantification of insulin: a comparison of an LC-MS/MS assay and immunoassays

Tue Søeborg, Márta Varga, J Sigh, Nikoline Juul Nielsen et al.
1 citations
Bioanalysis
Biosimilars and Bioanalytical Methods
0.24
article

Interference from anti-drug antibodies on the quantification of insulin: a comparison of an LC-MS/MS assay and immunoassays

Tue Søeborg, Márta Varga, J Sigh, Nikoline Juul Nielsen, Manca Spendal
article en
1 citations

Abstract

<b>Aim:</b> This study aims to compare the anti-drug antibody (ADA) interference in four pharmacokinetic (PK) assays across different platforms (AlphaLISA, Gyrolab, LC-MS/MS) and to devise a strategy for ADA interference mitigation to improve the accuracy of measured drug in total PK assays.<b>Materials & methods:</b> Spiked test samples, created to achieve different ADA concentrations in human serum also containing an insulin analogue, were analyzed alongside pooled clinical samples using four assays.<b>Results & conclusion:</b> Interference was observed in all platforms. A novel approach using the Gyrolab mixing CD, including acid dissociation in the PK assay, significantly reduced interference and thereby improved relative error from >99% to ≤20% yielding measurements well within the acceptance criteria. Clinical sample results reinforced findings from the test samples.

BioanalysisVol. 16(17-18)
University of Copenhagen (DK), Novo Nordisk (Denmark) (DK)
Good health and well-being
Openalex Percentile: Top 42%
Biosimilars and Bioanalytical Methods
0.24
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.