Cavity ionization chamber responses in the JAEA and the NMIJ high-energy photon reference fields for radiation protection
A dosimeter should ideally be calibrated in a reference field with similar energy and doserate to that which the dosimeter is being used to measure. Environments around nuclear reactors and radiation therapy facilities have high-energy photons with energies exceeding that of<sup>60</sup>Co gamma rays, and controlling exposure to these photons is important. The Japan Atomic Energy Agency and National Metrology Institute of Japan have high-energy reference fields with energies above several megaelectronvolts for different types of accelerators. Their reference fields have different fluence-energy distributions. In this study, the energy dependencies of the two-cavity ionization chambers, which are often used by secondary standard laboratories, were experimentally and computationally evaluated for each high-energy field. These results agreed well within the relative expanded uncertainties (<i>k</i>= 2), and their capabilities for air kerma measurements in each high-energy reference field were confirmed. Therefore, the capabilities of the air-kerma measurements were verified in the two high-energy reference fields.
Authors
- Tadahiro Kurosawa (ORCID: https://orcid.org/0000-0002-7909-2758)
- Morihito Shimizu (ORCID: https://orcid.org/0000-0002-3970-6735)
- Hiroshi Watabe (ORCID: https://orcid.org/0000-0001-6257-7735)
- Junya Ishii (ORCID: https://orcid.org/0000-0002-0400-3933)
- Hiroshi Yoshitomi (ORCID: https://orcid.org/0000-0003-1808-5615)
- Masahiro Kato (ORCID: https://orcid.org/0000-0003-4050-023X)
- Yoshihiko Tanimura (ORCID: https://orcid.org/0000-0003-4356-4765)
- Tomoya Tsuji
Institutions
- Japan Atomic Energy Agency (JP)
- Tohoku University (JP)
- Japan Radioisotope Association (JP)
- National Institute of Advanced Industrial Science and Technology (JP)
Publication Details
- Journal
- Journal of Radiological Protection
- Published
- 2024-09-01
- DOI
- https://doi.org/10.1088/1361-6498/ad7a07
- Primary Topic
- Radiation Therapy and Dosimetry
- Type
- article
- Field-Weighted Citation Impact
- 0.00