Demonstration of in-vivo simultaneous 3D imaging with 18F-FDG and Na131I using Compton–PET system

Simultaneous imaging of the SPECT tracer <sup>131</sup>I and PET tracer <sup>18</sup>F is important in the diagnosis of high- and low-grade thyroid cancers because high-grade thyroid cancers have high <sup>18</sup>F-FDG and low <sup>131</sup>I uptake, while low-grade thyroid cancers have high <sup>131</sup>I and low <sup>18</sup>F-FDG uptake. In this study, Na<sup>131</sup>I and <sup>18</sup>F-FDG were simultaneously imaged using the Compton-PET system, in vivo. The angular resolution and sensitivity of the Compton camera with 356 keV gamma ray measured using a <sup>133</sup>Ba point source were 12.3° and 2 × 10<sup>-5</sup>, respectively. The spatial resolution and sensitivity of PET were measured with a <sup>22</sup>Na point source. The transaxial and axial spatial resolutions of the PET at the center of the FOV were 1.15 mm and 2.04 mm, respectively. Its sensitivity was 1.2 × 10<sup>-4</sup>. In-vivo images of the <sup>18</sup>F and <sup>131</sup>I isotopes were simultaneously acquired from mice. These showed that <sup>18</sup>F-FDG was active in the heart, brown fat, and brain, while Na<sup>131</sup>I was active in the thyroid, stomach, and bladder. Artifacts were found in the Compton camera images when the activity of <sup>131</sup>I was much lower than that of <sup>18</sup>F. This study demonstrates the potential of simultaneous clinical imaging of <sup>18</sup>F and <sup>131</sup>I.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2024-09-09
DOI
https://doi.org/10.1038/s41598-024-71750-3
Citations
3
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
1.21

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Demonstration of in-vivo simultaneous 3D imaging with 18F-FDG and Na131I using Compton–PET system

Linlin Yan, Hiroyuki Takahashi, Mizuki Uenomachi, Masao Yoshino et al.
3 citations
Scientific Reports
Medical Imaging Techniques and Applications
1.21
article

Demonstration of in-vivo simultaneous 3D imaging with 18F-FDG and Na131I using Compton–PET system

Linlin Yan, Hiroyuki Takahashi, Mizuki Uenomachi, Masao Yoshino, Kei Kamada, Kenichiro Ogane, Kenji Shimazoe, Donghwan Kim
article en
3 citations

Abstract

Simultaneous imaging of the SPECT tracer <sup>131</sup>I and PET tracer <sup>18</sup>F is important in the diagnosis of high- and low-grade thyroid cancers because high-grade thyroid cancers have high <sup>18</sup>F-FDG and low <sup>131</sup>I uptake, while low-grade thyroid cancers have high <sup>131</sup>I and low <sup>18</sup>F-FDG uptake. In this study, Na<sup>131</sup>I and <sup>18</sup>F-FDG were simultaneously imaged using the Compton-PET system, in vivo. The angular resolution and sensitivity of the Compton camera with 356 keV gamma ray measured using a <sup>133</sup>Ba point source were 12.3° and 2 × 10<sup>-5</sup>, respectively. The spatial resolution and sensitivity of PET were measured with a <sup>22</sup>Na point source. The transaxial and axial spatial resolutions of the PET at the center of the FOV were 1.15 mm and 2.04 mm, respectively. Its sensitivity was 1.2 × 10<sup>-4</sup>. In-vivo images of the <sup>18</sup>F and <sup>131</sup>I isotopes were simultaneously acquired from mice. These showed that <sup>18</sup>F-FDG was active in the heart, brown fat, and brain, while Na<sup>131</sup>I was active in the thyroid, stomach, and bladder. Artifacts were found in the Compton camera images when the activity of <sup>131</sup>I was much lower than that of <sup>18</sup>F. This study demonstrates the potential of simultaneous clinical imaging of <sup>18</sup>F and <sup>131</sup>I.

Scientific ReportsVol. 14(1)
Bunkyo University (JP), Tohoku University (JP), Kyoto University (JP), Mita Hospital (JP), The University of Tokyo (JP), International University of Health and Welfare (JP)
Japan Society for the Promotion of Science
Good health and well-being
Openalex Percentile: Top 20%
Medical Imaging Techniques and Applications
1.21
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.