A comparative study of central corneal thickness using an ultrasonic pachymeter and a noncontact specular microscope in adult patients coming to the ophthalmology outpatient department in a tertiary care hospital in Western India

Objective: To evaluate and compare the accuracy of measuring central corneal thickness (CCT) using an ultrasonic pachymeter and a noncontact specular microscope and whether they can be used interchangeably in adult patients with healthy cornea. Methodology: A comparative and observational study was carried out in adult patients coming to the ophthalmology outpatient department (OPD) in a tertiary care hospital in Western India. In all patients, CCT is measured using an ultrasonic pachymeter and a noncontact specular microscope to perform a comparative study. Statistical analysis was done using paired t -test. Two-tailed hypothesis was applied. Results: This study included a sample size of 300 eyes of 150 adult patients. We want to test whether CCT values by Ultrasound Pachymeter and Noncontact specular Microscope is significantly different for right eyes and left eyes of the same patients, therefor we applied paired t -test for statistical analysis. T value for the right eye CCT measurement was −0.4 and the value of P was 0.69. T value for the left eye CCT measurement was −0.79 and the value of P was 0.43. P > 0.05 suggested that the results were nonsignificant. Conclusion: Our study suggests that the noncontact specular microscope produces CCT values that do not vary significantly from ultrasonic pachymeter. In subjects with healthy cornea, ultrasound pachymeter and specular microscope can be used interchangeably.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Ophthalmology and Research
Published
2024-09-01
DOI
https://doi.org/10.4103/jcor.jcor_67_24
Citations
1
Primary Topic
Corneal surgery and disorders
Type
article
Field-Weighted Citation Impact
0.40
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A comparative study of central corneal thickness using an ultrasonic pachymeter and a noncontact specular microscope in adult patients coming to the ophthalmology outpatient department in a tertiary care hospital in Western India

Ashwini A. Sapre, Krutika Nankani, Twisha M. Tandel, Hemali Arvindbhai Patel
1 citations
Journal of Clinical Ophthalmology and Research
Corneal surgery and disorders
0.40
article

A comparative study of central corneal thickness using an ultrasonic pachymeter and a noncontact specular microscope in adult patients coming to the ophthalmology outpatient department in a tertiary care hospital in Western India

Ashwini A. Sapre, Krutika Nankani, Twisha M. Tandel, Hemali Arvindbhai Patel
article en
1 citations

Abstract

Objective: To evaluate and compare the accuracy of measuring central corneal thickness (CCT) using an ultrasonic pachymeter and a noncontact specular microscope and whether they can be used interchangeably in adult patients with healthy cornea. Methodology: A comparative and observational study was carried out in adult patients coming to the ophthalmology outpatient department (OPD) in a tertiary care hospital in Western India. In all patients, CCT is measured using an ultrasonic pachymeter and a noncontact specular microscope to perform a comparative study. Statistical analysis was done using paired t -test. Two-tailed hypothesis was applied. Results: This study included a sample size of 300 eyes of 150 adult patients. We want to test whether CCT values by Ultrasound Pachymeter and Noncontact specular Microscope is significantly different for right eyes and left eyes of the same patients, therefor we applied paired t -test for statistical analysis. T value for the right eye CCT measurement was −0.4 and the value of P was 0.69. T value for the left eye CCT measurement was −0.79 and the value of P was 0.43. P > 0.05 suggested that the results were nonsignificant. Conclusion: Our study suggests that the noncontact specular microscope produces CCT values that do not vary significantly from ultrasonic pachymeter. In subjects with healthy cornea, ultrasound pachymeter and specular microscope can be used interchangeably.

Journal of Clinical Ophthalmology and ResearchVol. 12(3)
Sir Sayajirao General Hospital Medical College (IN)
Openalex Percentile: Top 35%
Corneal surgery and disorders
0.40
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

A comparative study of central corneal thickness using an ultrasonic pachymeter and a noncontact specular microscope in adult patients coming to the ophthalmology outpatient department in a tertiary care hospital in Western India — Ashwini A. Sapre, Krutika Nankani, et al. · Journal of Clinical Ophthalmology and Research (2024) | TGRS Research Map | TGRS