Peripheral Non-Contrast MR Angiography Using FBI: Scan Time and T2 Blurring Reduction with 2D Parallel Imaging

Non-contrast magnetic resonance angiography (NC-MRA), including fresh blood imaging (FBI), is a suitable choice for evaluating patients with peripheral artery disease (PAD). We evaluated standard FBI (sFBI) and centric ky-kz FBI (cFBI) acquisitions, using 1D and 2D parallel imaging factors (PIFs) to assess the trade-off between scan time and image quality due to blurring. The bilateral legs of four volunteers (mean age 33 years, two females) were imaged in the coronal plane using a body array coil with a posterior spine coil. Two types of sFBI and cFBI sequences with 1D PIF factor 5 in the phase encode (PE) direction (in-plane) and 2D PIF 3 (PE) × 2 (slice encode (SE)) (in-plane, through-slice) were studied. Image quality was evaluated by a radiologist, the vessel's signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were measured, and major vessel width was measured on the coronal maximum intensity projection (MIP) and 80-degree MIP. Results showed significant time reductions from 184 to 206 s on average when using sFBI down to 98 to 162 s when using cFBI (<i>p</i> = 0.003). Similar SNRs (averaging 200 to 370 across all sequences and PIF) and CNRs (averaging 190 to 360) for all techniques (<i>p</i> > 0.08) were found. There was no significant difference in the image quality (averaging 4.0 to 4.5; <i>p</i> > 0.2) or vessel width (averaging 4.1 to 4.9 mm; <i>p</i> > 0.1) on coronal MIP due to sequence or PIF. However, vessel width measured using 80-degree MIP demonstrated a significantly wider vessel in cFBI (5.6 to 6.8 mm) compared to sFBI (4.5 to 4.7 mm) (<i>p</i> = 0.022), and in 1D (4.7 to 6.8 mm) compared to 2D (4.5 to 5.6 mm) (<i>p</i> < 0.05) PIF. This demonstrated a trade-off in T2 blurring between 1D and 2D PIF: 1D using a PIF of 5 shortened the acquisition window, resulting in sharper arterial blood vessels in coronal images but significant blur in the 80-degree MIP. Two-dimensional PIF for cFBI provided a good balance between shorter scan time (relative to sFBI) and good sharpness in both in- and through-plane, while no benefit of 2D PIF was seen for sFBI. In conclusion, this study demonstrated the usefulness of FBI-based techniques for peripheral artery imaging and underscored the need to strike a balance between scan time and image quality in different planes through the use of 2D parallel imaging.

Authors

Institutions

Publication Details

Journal
Journal of Imaging
Published
2024-09-09
DOI
https://doi.org/10.3390/jimaging10090223
Citations
1
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.40

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Peripheral Non-Contrast MR Angiography Using FBI: Scan Time and T2 Blurring Reduction with 2D Parallel Imaging

Won C. Bae, Vadim Malis, Lewis D. Hahn, Diana Vucevic et al.
1 citations
Journal of Imaging
Advanced MRI Techniques and Applications
0.40
article

Peripheral Non-Contrast MR Angiography Using FBI: Scan Time and T2 Blurring Reduction with 2D Parallel Imaging

Won C. Bae, Vadim Malis, Lewis D. Hahn, Diana Vucevic, Mitsue Miyazaki, Anya Mesa
article en
1 citations

Abstract

Non-contrast magnetic resonance angiography (NC-MRA), including fresh blood imaging (FBI), is a suitable choice for evaluating patients with peripheral artery disease (PAD). We evaluated standard FBI (sFBI) and centric ky-kz FBI (cFBI) acquisitions, using 1D and 2D parallel imaging factors (PIFs) to assess the trade-off between scan time and image quality due to blurring. The bilateral legs of four volunteers (mean age 33 years, two females) were imaged in the coronal plane using a body array coil with a posterior spine coil. Two types of sFBI and cFBI sequences with 1D PIF factor 5 in the phase encode (PE) direction (in-plane) and 2D PIF 3 (PE) × 2 (slice encode (SE)) (in-plane, through-slice) were studied. Image quality was evaluated by a radiologist, the vessel's signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were measured, and major vessel width was measured on the coronal maximum intensity projection (MIP) and 80-degree MIP. Results showed significant time reductions from 184 to 206 s on average when using sFBI down to 98 to 162 s when using cFBI (<i>p</i> = 0.003). Similar SNRs (averaging 200 to 370 across all sequences and PIF) and CNRs (averaging 190 to 360) for all techniques (<i>p</i> > 0.08) were found. There was no significant difference in the image quality (averaging 4.0 to 4.5; <i>p</i> > 0.2) or vessel width (averaging 4.1 to 4.9 mm; <i>p</i> > 0.1) on coronal MIP due to sequence or PIF. However, vessel width measured using 80-degree MIP demonstrated a significantly wider vessel in cFBI (5.6 to 6.8 mm) compared to sFBI (4.5 to 4.7 mm) (<i>p</i> = 0.022), and in 1D (4.7 to 6.8 mm) compared to 2D (4.5 to 5.6 mm) (<i>p</i> < 0.05) PIF. This demonstrated a trade-off in T2 blurring between 1D and 2D PIF: 1D using a PIF of 5 shortened the acquisition window, resulting in sharper arterial blood vessels in coronal images but significant blur in the 80-degree MIP. Two-dimensional PIF for cFBI provided a good balance between shorter scan time (relative to sFBI) and good sharpness in both in- and through-plane, while no benefit of 2D PIF was seen for sFBI. In conclusion, this study demonstrated the usefulness of FBI-based techniques for peripheral artery imaging and underscored the need to strike a balance between scan time and image quality in different planes through the use of 2D parallel imaging.

Journal of ImagingVol. 10(9)
University of California San Diego (US), VA San Diego Healthcare System (US)
Canon Medical Systems USA
Good health and well-being
Openalex Percentile: Top 35%
Advanced MRI Techniques and Applications
0.40
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.