EP12.12: Agreement between two different devices for maternal hemodynamic evaluation

This study aimed to assess the agreement between maternal hemodynamic measurements taken with two different devices for maternal hemodynamic evaluation. Overall, 344 measurements from 172 patients (156 healthy controls, 11 patients with small for gestational age fetuses and 5 with fetal growth restriction) enrolled at the Imperial College, London were considered. The agreement between USCOM® and Vicorder® measurements was evaluated for Heart Rate (HR), Cardiac Output (CO), Stroke Volume (SV) and Total Peripheral Resistance (TPR). The assessment was based on Bland-Altman plots with 95% limits of agreement (LOA) for the mean difference (MD) between two devices, reporting also the intraclass correlation coefficient (ICC). In general, measures from different devices were poorly correlated, except for HR (ICC = 0.76, p < 0.001). In particular, ICC was 0.21 for TPR, 0.28 for SV and 0.29 for CO. Wide LOA were found for all measurements. USCOM® provided, on average, higher values for HR (MD = -0.30), CO (MD = -0.47) and SV (MD = -5.13) and lower for TPR (MD = 72.7), compared to Vicorder®. The agreement between USCOM® and Vicorder® measurements is low, emphasising the high risk of bias in using measurements from different devices without standardisation. Thus, when data from different devices are compared, standardisation based on reference ranges should be considered (i.e. z-scores, modified z-scores, MoMs or similar). Please note: The publisher is not responsible for the content or functionality of any supporting information supplied by the authors. Any queries (other than missing content) should be directed to the corresponding author for the article.

Authors

Institutions

Publication Details

Journal
Ultrasound in Obstetrics and Gynecology
Published
2024-09-01
DOI
https://doi.org/10.1002/uog.28564
Primary Topic
Hemodynamic Monitoring and Therapy
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

EP12.12: Agreement between two different devices for maternal hemodynamic evaluation

Giulia Zamagni, Roberta Erra, Lorenzo Monasta, Tamara Stampalija et al.
Ultrasound in Obstetrics and Gynecology
Hemodynamic Monitoring and Therapy
article

EP12.12: Agreement between two different devices for maternal hemodynamic evaluation

Giulia Zamagni, Roberta Erra, Lorenzo Monasta, Tamara Stampalija, Bronacha Mylrea‐Foley, C. Lees, John L. Goodier, L. Jones
article en

Abstract

This study aimed to assess the agreement between maternal hemodynamic measurements taken with two different devices for maternal hemodynamic evaluation. Overall, 344 measurements from 172 patients (156 healthy controls, 11 patients with small for gestational age fetuses and 5 with fetal growth restriction) enrolled at the Imperial College, London were considered. The agreement between USCOM® and Vicorder® measurements was evaluated for Heart Rate (HR), Cardiac Output (CO), Stroke Volume (SV) and Total Peripheral Resistance (TPR). The assessment was based on Bland-Altman plots with 95% limits of agreement (LOA) for the mean difference (MD) between two devices, reporting also the intraclass correlation coefficient (ICC). In general, measures from different devices were poorly correlated, except for HR (ICC = 0.76, p < 0.001). In particular, ICC was 0.21 for TPR, 0.28 for SV and 0.29 for CO. Wide LOA were found for all measurements. USCOM® provided, on average, higher values for HR (MD = -0.30), CO (MD = -0.47) and SV (MD = -5.13) and lower for TPR (MD = 72.7), compared to Vicorder®. The agreement between USCOM® and Vicorder® measurements is low, emphasising the high risk of bias in using measurements from different devices without standardisation. Thus, when data from different devices are compared, standardisation based on reference ranges should be considered (i.e. z-scores, modified z-scores, MoMs or similar). Please note: The publisher is not responsible for the content or functionality of any supporting information supplied by the authors. Any queries (other than missing content) should be directed to the corresponding author for the article.

Ultrasound in Obstetrics and GynecologyVol. 64(S1)
University of Trieste (IT), Imperial College Healthcare NHS Trust (GB), IRCCS Materno Infantile Burlo Garofolo (IT), Imperial College London (GB)
Zero hunger
Openalex Percentile: Top 72%
Hemodynamic Monitoring and Therapy
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.