OP08.04: Mechanical properties of cyanoacrylate‐based bioadhesives and Nitinol meshes to develop a sealing system to prevent iPPROM

To evaluate the mechanical properties of different cyanoacrylate-based bioadhesives and Nitinol meshes for improving a sealing system for fetoscopy-induced membrane defects. Two-stage ex vivo experiments were conducted to evaluate different types of bioadhesives and Nitinol meshes. The first stage involved selecting the ideal bioadhesive among n-cyanoacrylate (nCA)-viscosity-inducing agent (VIA) 5%, nCA-VIA 6% and nCA-VIA 7%. The second stage involved selecting the ideal Nitinol mesh design(s) among silicone-Nitinol-6-arm-thin, silicone-Nitinol-6-arm-thick, silicone-Nitinol-8-arm-thin and silicone-Nitinol-8-arm-thick patches. Term human amniotic membranes were punctured with a 10F metal punch, and the defects were sealed with the sealing patches. The membranes were mounted on a multiaxial inflation device, and fluorescein saline solution was infused to generate hydrostatic pressure that mimics intraamniotic pressure. The maximum pressure free of leakage (MP) and bursting pressure (BP) were measured (n = 8). Intact membranes were used as controls for the two-stage experiments. There was no significant difference in MP and BP among the three bioadhesives. However, nCA-VIA 7% showed a more consistent performance in all experiments and smaller difference between MP and BP, compared with others. In the experiments for Nitinol mesh selection, silicone-Nitinol-6-arm-thin patches had a significantly higher MP than silicone-Nitinol-8-arm-thin patches (P = 0.018). Silicone-Nitinol-6-arm-thin and silicone-Nitinol-6-arm-thick patches showed higher BP, compared with other patches, with a trend towards statistical significance (P = 0.059). nCA-VIA 7% and silicone-Nitinol-6-arm-thin and silicone-Nitinol-6-arm-thick patches showed the best performance in terms of adhesion and mechanical properties. Further experiments to evaluate intermediate and long-term adhesion are needed to select the best candidates for sealing system development.

Authors

Institutions

Publication Details

Journal
Ultrasound in Obstetrics and Gynecology
Published
2024-09-01
DOI
https://doi.org/10.1002/uog.27936
Primary Topic
Surgical Sutures and Adhesives
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

OP08.04: Mechanical properties of cyanoacrylate‐based bioadhesives and Nitinol meshes to develop a sealing system to prevent iPPROM

Salvador Borrós, T.M. Helfer, Eduard Gratacós, G. Febas et al.
Ultrasound in Obstetrics and Gynecology
Surgical Sutures and Adhesives
article

OP08.04: Mechanical properties of cyanoacrylate‐based bioadhesives and Nitinol meshes to develop a sealing system to prevent iPPROM

Salvador Borrós, T.M. Helfer, Eduard Gratacós, G. Febas, E. Eixarch, Sun Yong, M. Gómez Herranz
article en

Abstract

To evaluate the mechanical properties of different cyanoacrylate-based bioadhesives and Nitinol meshes for improving a sealing system for fetoscopy-induced membrane defects. Two-stage ex vivo experiments were conducted to evaluate different types of bioadhesives and Nitinol meshes. The first stage involved selecting the ideal bioadhesive among n-cyanoacrylate (nCA)-viscosity-inducing agent (VIA) 5%, nCA-VIA 6% and nCA-VIA 7%. The second stage involved selecting the ideal Nitinol mesh design(s) among silicone-Nitinol-6-arm-thin, silicone-Nitinol-6-arm-thick, silicone-Nitinol-8-arm-thin and silicone-Nitinol-8-arm-thick patches. Term human amniotic membranes were punctured with a 10F metal punch, and the defects were sealed with the sealing patches. The membranes were mounted on a multiaxial inflation device, and fluorescein saline solution was infused to generate hydrostatic pressure that mimics intraamniotic pressure. The maximum pressure free of leakage (MP) and bursting pressure (BP) were measured (n = 8). Intact membranes were used as controls for the two-stage experiments. There was no significant difference in MP and BP among the three bioadhesives. However, nCA-VIA 7% showed a more consistent performance in all experiments and smaller difference between MP and BP, compared with others. In the experiments for Nitinol mesh selection, silicone-Nitinol-6-arm-thin patches had a significantly higher MP than silicone-Nitinol-8-arm-thin patches (P = 0.018). Silicone-Nitinol-6-arm-thin and silicone-Nitinol-6-arm-thick patches showed higher BP, compared with other patches, with a trend towards statistical significance (P = 0.059). nCA-VIA 7% and silicone-Nitinol-6-arm-thin and silicone-Nitinol-6-arm-thick patches showed the best performance in terms of adhesion and mechanical properties. Further experiments to evaluate intermediate and long-term adhesion are needed to select the best candidates for sealing system development.

Ultrasound in Obstetrics and GynecologyVol. 64(S1)
Hospital Sant Joan de Déu Barcelona (ES), Centre for Biomedical Network Research on Rare Diseases (ES), Institut Químic de Sarrià (ES), Universitat Ramon Llull (ES), Universitat de Barcelona (ES)
Openalex Percentile: Top 72%
Surgical Sutures and Adhesives
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.