An efficient component of the redundancy calibration program to ensure equipment stability by assaying HDR Ir‐192 sources at the time of replacement
The method proposed allows any hospital to ensure a redundancy component of the long-term stability of all equipment involved in BT measurements in a very simple and time efficient manner. Additionally, it enables the hospital to maintain a detailed log of historical differences, facilitating the identification and correction of potential systematic deviations over time.