The concurrent validity of a portable ultrasound probe for muscle thickness measurements

Ultrasound imaging is extensively used by both practitioners and researchers in assessing muscle thickness (MT); however, its use in the field is constrained by the transportability of stationary devices. New portable ultrasound probes pose as a cost-effective and transportable alternative for field-based assessments. This study evaluated the concurrent validity of a portable probe (Lumify) against a laboratory-based device (Vivid S5) in measuring MT. Eighteen participants (nine males and nine females) visited the laboratory and their MT measurements were collected using each device at five different sites (anterior and posterior arm, anterior and posterior thigh, and posterior lower leg). Bland-Altman plots (systematic and proportional bias, random error, and 95% limits of agreement), Pearson's product-moment correlation coefficient (r), and paired samples t-tests with Cohen's d effect sizes (ES) were used to assess the concurrent validity of the Lumify device. Systematic bias was low at all sites ( ≤ 0.11 cm) while proportional bias was detected only at the posterior lower leg (r<sup>2</sup> = 0.217 [r = 0.466]). The difference in MT between devices was significant only at the anterior thigh (p < 0.05); however, ES for all sites were considered trivial (ES ≤ 0.131). Linear associations were found between the devices at each site of measurement (r ≥ 0.95). These results highlight that the Lumify probe can be used interchangeably with the Vivid S5 for MT measurements, providing practitioners and researchers with a more cost-effective and portable alternative for field-based assessments.

Authors

Institutions

Publication Details

Journal
Clinical Physiology and Functional Imaging
Published
2024-09-05
DOI
https://doi.org/10.1111/cpf.12901
Citations
1
Primary Topic
Body Composition Measurement Techniques
Type
article
Field-Weighted Citation Impact
0.33

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The concurrent validity of a portable ultrasound probe for muscle thickness measurements

Ivan Jukić, Matt R. Cross, Kai A. Homer
1 citations
Clinical Physiology and Functional Imaging
Body Composition Measurement Techniques
0.33
article

The concurrent validity of a portable ultrasound probe for muscle thickness measurements

Ivan Jukić, Matt R. Cross, Kai A. Homer
article en
1 citations

Abstract

Ultrasound imaging is extensively used by both practitioners and researchers in assessing muscle thickness (MT); however, its use in the field is constrained by the transportability of stationary devices. New portable ultrasound probes pose as a cost-effective and transportable alternative for field-based assessments. This study evaluated the concurrent validity of a portable probe (Lumify) against a laboratory-based device (Vivid S5) in measuring MT. Eighteen participants (nine males and nine females) visited the laboratory and their MT measurements were collected using each device at five different sites (anterior and posterior arm, anterior and posterior thigh, and posterior lower leg). Bland-Altman plots (systematic and proportional bias, random error, and 95% limits of agreement), Pearson's product-moment correlation coefficient (r), and paired samples t-tests with Cohen's d effect sizes (ES) were used to assess the concurrent validity of the Lumify device. Systematic bias was low at all sites ( ≤ 0.11 cm) while proportional bias was detected only at the posterior lower leg (r<sup>2</sup> = 0.217 [r = 0.466]). The difference in MT between devices was significant only at the anterior thigh (p < 0.05); however, ES for all sites were considered trivial (ES ≤ 0.131). Linear associations were found between the devices at each site of measurement (r ≥ 0.95). These results highlight that the Lumify probe can be used interchangeably with the Vivid S5 for MT measurements, providing practitioners and researchers with a more cost-effective and portable alternative for field-based assessments.

Clinical Physiology and Functional ImagingVol. 45(1)
Auckland University of Technology (NZ), Abertay University (GB)
Auckland University of Technology, New Zealand
Openalex Percentile: Top 38%
Body Composition Measurement Techniques
0.33
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.