Timepix4 characterization with monochromatic X-rays at the Elettra synchrotron facility

Timepix4 is the latest generation application-specific integrated circuit (ASIC) of the Medipix family, developed by the Medipix4 Collaboration. A characterization of a Timepix4 assembly bump-bonded to a planar 300 μ m thick silicon pixelated sensor has been performed by using monochromatic X-rays at the Elettra synchrotron facility (Trieste, Italy). In particular, a calibration of the ToT against the energy released in the Si detector has been performed for each pixel, to optimize energy resolution. A preliminary analysis to estimate the detector photon counting linearity and pixel readout dead time was also carried out.

Authors

Institutions

Publication Details

Journal
Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment
Published
2024-09-06
DOI
https://doi.org/10.1016/j.nima.2024.169844
Citations
7
Primary Topic
Particle Detector Development and Performance
Type
article
Field-Weighted Citation Impact
4.62

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Timepix4 characterization with monochromatic X-rays at the Elettra synchrotron facility

Francesco Brun, P. Cardarelli, V. Cavallini, N. V. Biesuz et al.
7 citations
Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment
Particle Detector Development and Performance
4.62
article

Timepix4 characterization with monochromatic X-rays at the Elettra synchrotron facility

Francesco Brun, P. Cardarelli, V. Cavallini, N. V. Biesuz, R. Bolzonella, Luca Brombal, A. Feruglio
article en
7 citations

Abstract

Timepix4 is the latest generation application-specific integrated circuit (ASIC) of the Medipix family, developed by the Medipix4 Collaboration. A characterization of a Timepix4 assembly bump-bonded to a planar 300 μ m thick silicon pixelated sensor has been performed by using monochromatic X-rays at the Elettra synchrotron facility (Trieste, Italy). In particular, a calibration of the ToT against the energy released in the Si detector has been performed for each pixel, to optimize energy resolution. A preliminary analysis to estimate the detector photon counting linearity and pixel readout dead time was also carried out.

Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated EquipmentVol. 1069
University of Trieste (IT), University of Ferrara (IT), Istituto Nazionale di Fisica Nucleare, Sezione di Ferrara (IT), Istituto Nazionale di Fisica Nucleare, Sezione di Trieste (IT), Istituto Nazionale di Fisica Nucleare, Sezione di Pisa (IT)
Commissione Scientifica Nazionale 5, Instituto Nazionale di Fisica Nucleare
Affordable and clean energy
Openalex Percentile: Top 4%
Particle Detector Development and Performance
4.62
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.