Effect of interface dead space on the time taken to achieve changes in set FiO<sub>2</sub> during T-piece ventilation: is face mask the optimal interface for neonatal stabilisation?
There was a delay in achieving changes in oxygen delivery at the airway during simulated ventilation attributable to the mask volume. This delay was greatly reduced when using nasal prongs as an interface. This should be examined in clinical trials.