Evidence of time dependent degradation of polypropylene surgical mesh explanted from the abdomen and vagina of sheep

The failure of polypropylene mesh is marked by significant side effects and debilitation, arising from a complex interplay of factors. One key contributor is the pronounced physico-mechanical mismatch between the polypropylene (PP) fibres and surrounding tissues, resulting in substantial physical damage, inflammation, and persistent pain. However, the primary cause of sustained inflammation due to polypropylene itself remains incompletely understood. This study comprises a comprehensive, multi-pronged investigation to unravel the effects of implantation on a presumed inert PP mesh in sheep. Employing both advanced and conventional techniques to discern the physical and chemical transformations of the implanted PP. Our analyses reveal a surface degradation and oxidation of polypropylene fibres after 60 days implantation, persisting and intensifying at the 180-day mark. The emergence and accumulation of PP debris in the tissue surrounding the implant also increased with implantation time. We demonstrate observable physical and mechanical alterations in the fibre surface and stiffness. Our study shows surface alterations which indicate that PP is evidently less chemically inert than was initially presumed. These findings underscore the need for a re-evaluation of the biocompatibility and long-term consequences of using PP mesh implants.

Authors

Institutions

Publication Details

Journal
Journal of the mechanical behavior of biomedical materials/Journal of mechanical behavior of biomedical materials
Published
2024-09-06
DOI
https://doi.org/10.1016/j.jmbbm.2024.106722
Citations
12
Primary Topic
Hernia repair and management
Type
article
Field-Weighted Citation Impact
6.87

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evidence of time dependent degradation of polypropylene surgical mesh explanted from the abdomen and vagina of sheep

Cassandra Rauert, Cornelia Rodenburg, Alexander J. Knight, Kevin Victor Thomas et al.
12 citations
Journal of the mechanical behavior of biomedical materials/Journal of mechanical behavior of biomedical materials
Hernia repair and management
6.87
article

Evidence of time dependent degradation of polypropylene surgical mesh explanted from the abdomen and vagina of sheep

Cassandra Rauert, Cornelia Rodenburg, Alexander J. Knight, Kevin Victor Thomas, David A. Gregory, Sabiniano Román, Jan A. Deprest, Christopher Reginald Chapple, Victoria L. Workman, Nicholas T. H. Farr, A. I. Tartakovskii, Anthony J. Bullock, Sheila MacNeil
article en
12 citations

Abstract

The failure of polypropylene mesh is marked by significant side effects and debilitation, arising from a complex interplay of factors. One key contributor is the pronounced physico-mechanical mismatch between the polypropylene (PP) fibres and surrounding tissues, resulting in substantial physical damage, inflammation, and persistent pain. However, the primary cause of sustained inflammation due to polypropylene itself remains incompletely understood. This study comprises a comprehensive, multi-pronged investigation to unravel the effects of implantation on a presumed inert PP mesh in sheep. Employing both advanced and conventional techniques to discern the physical and chemical transformations of the implanted PP. Our analyses reveal a surface degradation and oxidation of polypropylene fibres after 60 days implantation, persisting and intensifying at the 180-day mark. The emergence and accumulation of PP debris in the tissue surrounding the implant also increased with implantation time. We demonstrate observable physical and mechanical alterations in the fibre surface and stiffness. Our study shows surface alterations which indicate that PP is evidently less chemically inert than was initially presumed. These findings underscore the need for a re-evaluation of the biocompatibility and long-term consequences of using PP mesh implants.

Journal of the mechanical behavior of biomedical materials/Journal of mechanical behavior of biomedical materialsVol. 160
The University of Queensland (AU), Royal Hallamshire Hospital (GB), Sheffield Teaching Hospitals NHS Foundation Trust (GB), Sheffield Children's NHS Foundation Trust (GB), University of Sheffield (GB), KU Leuven (BE)
Diamond Light Source, Minderoo Foundation, Medical Research Council, Engineering and Physical Sciences Research Council
Zero hunger
Openalex Percentile: Top 2%
Hernia repair and management
6.87
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.