Performance evaluation of single‐ and dual‐contrast spectral imaging on a photon‐counting‐detector CT
Optimal tube voltage for material decomposition on the clinical PCD-CT is task-dependent but inferior to DS-DECT using 70/Sn150 kV or 80/Sn150 kV in two-material decomposition for single-contrast imaging (iodine/water or gadolinium/water). Three material decomposition (iodine/gadolinium/water) in dual-contrast imaging yields substantially higher RMSE for both imaging platforms.