Consistency in the Assessment of Dried Blood Spot Specimen Size and Quality in U.K. Newborn Screening Laboratories

In 2015, U.K. newborn screening (NBS) laboratory guidelines were introduced to standardize dried blood spot (DBS) specimen quality acceptance and specify a minimum acceptable DBS diameter of ≥7 mm. The UK 'acceptable' avoidable repeat rate (AVRR) is ≤2%. To assess inter-laboratory variability in specimen acceptance/rejection, two sets of colored scanned images (<i>n</i> = 40/set) of both good and poor-quality DBS specimens were distributed to all 16 U.K. NBS laboratories for evaluation as part of an external quality assurance (EQA) assessment. The mean (range) number of specimens rejected in the first EQA distribution was 7 (1-16) and in the second EQA distribution was 7 (0-16), demonstrating that adherence to the 2015 guidelines was highly variable. A new minimum standard for DBS size of ≥8 mm (to enable a minimum of six sub-punches from two DBS) was discussed. NBS laboratories undertook a prospective audit and demonstrated that using ≥8 mm as the minimum acceptable DBS diameter would increase the AVRR from 2.1% (range 0.55% to 5.5%) to 7.8% (range 0.55% to 22.7%). A significant inverse association between the number of specimens rejected in the DBS EQA distributions and the predicted AVVR (using ≥8 mm minimum standard) was observed (r = -0.734, <i>p</i> = 0.003). Before implementing more stringent standards, the impact of a standard operating procedure (SOP) designed to enable a standardized approach of visual assessment and using the existing ≥7 mm diameter (to enable a minimum of four sub-punches from two DBS) as the minimum standard was assessed in a retrospective audit. Implementation of the SOP and using the ≥7 mm DBS diameter would increase the AVRR from 2.3% (range 0.63% to 5.3%) to 6.5% (range 4.3% to 20.9%). The results demonstrate that there is inconsistency in applying the acceptance/rejection criteria, and that a low AVVR is not an indication of good-quality specimens being received into laboratories. Further work is underway to introduce and maintain standards without increasing the AVRR to unacceptable levels.

Authors

Institutions

Publication Details

Journal
International Journal of Neonatal Screening
Published
2024-09-05
DOI
https://doi.org/10.3390/ijns10030060
Citations
4
Primary Topic
Metabolism and Genetic Disorders
Type
article
Field-Weighted Citation Impact
1.52
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Consistency in the Assessment of Dried Blood Spot Specimen Size and Quality in U.K. Newborn Screening Laboratories

Christine Cavanagh, Beverly Hird, Claire Manfredonia, Tejswurree Ramgoolam et al.
4 citations
International Journal of Neonatal Screening
Metabolism and Genetic Disorders
1.52
article

Consistency in the Assessment of Dried Blood Spot Specimen Size and Quality in U.K. Newborn Screening Laboratories

Christine Cavanagh, Beverly Hird, Claire Manfredonia, Tejswurree Ramgoolam, Jennifer Cundick, Lynette Shakespeare, Stuart J. Moat, David Elliman, Sarah A. Smith, Catherine Collingwood, Caroline Griffith, Philippa Goddard, Rachel S. Carling, Clare Le Masurier, Nazia Taj, Mehdi Mirzazadeh, Nick Flynn, Tessa Morgan, James R. Bonham, Lesley Tetlow, Laura Wainwright, Fiona Jenkinson, L.E. Robinson, Margaret Birch, Amy Headley
article en
4 citations

Abstract

In 2015, U.K. newborn screening (NBS) laboratory guidelines were introduced to standardize dried blood spot (DBS) specimen quality acceptance and specify a minimum acceptable DBS diameter of ≥7 mm. The UK 'acceptable' avoidable repeat rate (AVRR) is ≤2%. To assess inter-laboratory variability in specimen acceptance/rejection, two sets of colored scanned images (<i>n</i> = 40/set) of both good and poor-quality DBS specimens were distributed to all 16 U.K. NBS laboratories for evaluation as part of an external quality assurance (EQA) assessment. The mean (range) number of specimens rejected in the first EQA distribution was 7 (1-16) and in the second EQA distribution was 7 (0-16), demonstrating that adherence to the 2015 guidelines was highly variable. A new minimum standard for DBS size of ≥8 mm (to enable a minimum of six sub-punches from two DBS) was discussed. NBS laboratories undertook a prospective audit and demonstrated that using ≥8 mm as the minimum acceptable DBS diameter would increase the AVRR from 2.1% (range 0.55% to 5.5%) to 7.8% (range 0.55% to 22.7%). A significant inverse association between the number of specimens rejected in the DBS EQA distributions and the predicted AVVR (using ≥8 mm minimum standard) was observed (r = -0.734, <i>p</i> = 0.003). Before implementing more stringent standards, the impact of a standard operating procedure (SOP) designed to enable a standardized approach of visual assessment and using the existing ≥7 mm diameter (to enable a minimum of four sub-punches from two DBS) as the minimum standard was assessed in a retrospective audit. Implementation of the SOP and using the ≥7 mm DBS diameter would increase the AVRR from 2.3% (range 0.63% to 5.3%) to 6.5% (range 4.3% to 20.9%). The results demonstrate that there is inconsistency in applying the acceptance/rejection criteria, and that a low AVVR is not an indication of good-quality specimens being received into laboratories. Further work is underway to introduce and maintain standards without increasing the AVRR to unacceptable levels.

International Journal of Neonatal ScreeningVol. 10(3)
King's College London (GB), Great Ormond Street Hospital (GB), Royal Victoria Infirmary (GB), University Hospital of Wales (GB), Southmead Hospital (GB), St James's University Hospital (GB), St Helier Hospital (GB), Cambridge University Hospitals NHS Foundation Trust (GB), Public Health Wales (GB), Portsmouth Hospitals NHS Trust (GB), Manchester University NHS Foundation Trust (GB), Alder Hey Children's NHS Foundation Trust (GB), Birmingham Children's Hospital (GB), Sheffield Children's NHS Foundation Trust (GB), Oxford University Hospitals NHS Trust (GB), Queen Elizabeth University Hospital (GB), Royal Victoria Hospital (GB), University College London (GB), Cardiff University (GB)
Openalex Percentile: Top 18%
Metabolism and Genetic Disorders
1.52
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.