This study compares the performance of old and new lung counters in the National Institutes for Quantum Sciences and Technology of Japan. The total sensitive area of the detector crystals for the new lung counter is ~15% smaller than that for the old lung counter. Minimum detectable activities (MDAs) for 241Am and 239Pu were evaluated through experiments using a Lawrence Livermore National Laboratory torso phantom. Despite differences in detector configuration, the MDAs were found to be comparable between the two lung counters. For a chest wall thickness of 2.1 cm and a counting time of 30 min, the MDAs of 241Am and 239Pu were 5.7 and 2300 Bq for the old lung counter, and 5.5 and 2600 Bq for the new lung counter, respectively. Experimental results for the relative sensitivities between left-side and right-side detectors suggested that the new lung counter offered better measurement geometry.

Authors

Institutions

Publication Details

Journal
Radiation Protection Dosimetry
Published
2024-09-01
DOI
https://doi.org/10.1093/rpd/ncae183
Citations
2
Primary Topic
Radiation Detection and Scintillator Technologies
Type
article
Field-Weighted Citation Impact
1.22
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A performance comparison study between two HPGe lung counters with different detector configurations

Munehiko Kowatari, Osamu Kurihara, E Kim, Kazuaki Yajima et al.
2 citations
Radiation Protection Dosimetry
Radiation Detection and Scintillator Technologies
1.22
article

A performance comparison study between two HPGe lung counters with different detector configurations

Munehiko Kowatari, Osamu Kurihara, E Kim, Kazuaki Yajima, Masayuki Naito, Kotaro Tani, Yuki Tamakuma, Yuma Mihei
article en
2 citations

Abstract

This study compares the performance of old and new lung counters in the National Institutes for Quantum Sciences and Technology of Japan. The total sensitive area of the detector crystals for the new lung counter is ~15% smaller than that for the old lung counter. Minimum detectable activities (MDAs) for 241Am and 239Pu were evaluated through experiments using a Lawrence Livermore National Laboratory torso phantom. Despite differences in detector configuration, the MDAs were found to be comparable between the two lung counters. For a chest wall thickness of 2.1 cm and a counting time of 30 min, the MDAs of 241Am and 239Pu were 5.7 and 2300 Bq for the old lung counter, and 5.5 and 2600 Bq for the new lung counter, respectively. Experimental results for the relative sensitivities between left-side and right-side detectors suggested that the new lung counter offered better measurement geometry.

Radiation Protection DosimetryVol. 200(15)
National Institutes for Quantum Science and Technology (JP), Nagasaki University (JP), Tokyo Metropolitan University (JP)
Openalex Percentile: Top 21%
Radiation Detection and Scintillator Technologies
1.22
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.