An optimized method for PFAS analysis using HR–CS–GFMAS via GaF detection

The analysis of per- and polyfluoroalkyl substances (PFAS) via sum parameters like extractable organic fluorine (EOF) in combination with high resolution-continuum source-graphite furnace molecular absorption spectrometry (HR-CS-GFMAS) is highly promising regarding fluorine sensitivity and selectivity. However, the HR-CS-GFMAS method includes several drying and heating steps which can lead to losses of volatile PFAS before the molecular formation step using e.g., GaF formation. Hence, the method leads to a strong discrimination of PFAS within the EOF depending on their physical/chemical properties and is therefore associated with reduced accuracy. To reduce this discrepancy and to indicate realistic PFAS pollution values, an optimization of the HR-CS-GFMAS method for PFAS analysis is needed. Hence, we determined fluorine response factors of several PFAS with different physical/chemical properties upon application of systematic optimization steps. We could therefore improve the method's sensitivity for PFAS analysis using a modifier drying pre-treatment step followed by a sequential injection of sample solutions. The highest improvement in sensitivity of volatile PFAS was shown upon addition of a Mg modifier during drying pre-treatment. Thereby, during optimization the relative standard deviation of fluorine response factors could be reduced from 55 % (initial method) to 27 % (optimized method) leading to a more accurate determination of organofluorine sum parameters. The method provides an instrumental LOD and LOQ of β(F) 1.71 μg/L and 5.13 μg/L, respectively. Further validation aimed to investigate several matrix effects with respect to water matrices. Here, substance-specific behavior was observed. For example, perfluorooctanoic acid (PFOA) which was used as calibrator, showed signal suppressions upon high chloride concentrations (>50 mg/L). Hence, a thorough separation of Cl from analytes during sample preparation is needed for accurate sum parameter analysis.

Authors

Institutions

Publication Details

Journal
Talanta
Published
2024-09-04
DOI
https://doi.org/10.1016/j.talanta.2024.126811
Citations
9
Primary Topic
Per- and polyfluoroalkyl substances research
Type
article
Field-Weighted Citation Impact
2.25

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

An optimized method for PFAS analysis using HR–CS–GFMAS via GaF detection

Lennart Gehrenkemper, Marcus von der Au, Fabian Simon, Björn Meermann
9 citations
Talanta
Per- and polyfluoroalkyl substances research
2.25
article

An optimized method for PFAS analysis using HR–CS–GFMAS via GaF detection

Lennart Gehrenkemper, Marcus von der Au, Fabian Simon, Björn Meermann
article en
9 citations

Abstract

The analysis of per- and polyfluoroalkyl substances (PFAS) via sum parameters like extractable organic fluorine (EOF) in combination with high resolution-continuum source-graphite furnace molecular absorption spectrometry (HR-CS-GFMAS) is highly promising regarding fluorine sensitivity and selectivity. However, the HR-CS-GFMAS method includes several drying and heating steps which can lead to losses of volatile PFAS before the molecular formation step using e.g., GaF formation. Hence, the method leads to a strong discrimination of PFAS within the EOF depending on their physical/chemical properties and is therefore associated with reduced accuracy. To reduce this discrepancy and to indicate realistic PFAS pollution values, an optimization of the HR-CS-GFMAS method for PFAS analysis is needed. Hence, we determined fluorine response factors of several PFAS with different physical/chemical properties upon application of systematic optimization steps. We could therefore improve the method's sensitivity for PFAS analysis using a modifier drying pre-treatment step followed by a sequential injection of sample solutions. The highest improvement in sensitivity of volatile PFAS was shown upon addition of a Mg modifier during drying pre-treatment. Thereby, during optimization the relative standard deviation of fluorine response factors could be reduced from 55 % (initial method) to 27 % (optimized method) leading to a more accurate determination of organofluorine sum parameters. The method provides an instrumental LOD and LOQ of β(F) 1.71 μg/L and 5.13 μg/L, respectively. Further validation aimed to investigate several matrix effects with respect to water matrices. Here, substance-specific behavior was observed. For example, perfluorooctanoic acid (PFOA) which was used as calibrator, showed signal suppressions upon high chloride concentrations (>50 mg/L). Hence, a thorough separation of Cl from analytes during sample preparation is needed for accurate sum parameter analysis.

TalantaVol. 281
Federal Institute For Materials Research and Testing (DE)
Deutsche Forschungsgemeinschaft
Openalex Percentile: Top 13%
Per- and polyfluoroalkyl substances research
2.25
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.