<h3>Introduction</h3> The OPTIQ, an advanced 2D imaging pipeline, automatically adjusts exposure parameters to maintain predefined image quality during cerebral angiography while aiming to reduce radiation dose. <h3>Aim of Study</h3> This study evaluates the efficacy of OPTIQ in diminishing radiation exposure without compromising image quality in low-dose diagnostic cerebral angiography. <h3>Methods</h3> We conducted a retrospective analysis of 2D diagnostic cerebral angiography data from 100 patients (68 female, 32 male; median age 58, range 31–80 years) assessed for intracranial aneurysms. Image quality was compared between 50 patients using conventional settings and 50 using OPTIQ with an Image Quality level of 36 (IQ36), evaluated by two blinded neurointerventionists on a four-point scale across four regions. Noninferiority in image quality and reductions in radiation dose were statistically analyzed using dose-area product (DAP) and air kerma (AK) measurements. <h3>Results</h3> The OPTIQ group demonstrated noninferior image quality (Reader 1: 3.50 ± 0.38; Reader 2: 3.70 ± 0.26) compared to the conventional group (Reader 1: 3.53 ± 0.37; Reader 2: 3.78 ± 0.19). Interrater reliability was established with a Cronbach’s alpha of 0.835. DAP and AK for 2D DSA in anteroposterior and lateral views were significantly lower in the OPTIQ group (1.94 Gy-cm² and 0.012 Gy, respectively) compared to the conventional group (5.28 Gy-cm² and 0.031 Gy, respectively), marking a dose reduction of 63.26% for DAP and 60.25% for AK. <h3>Conclusion</h3> OPTIQ with IQ36 significantly reduces radiation exposure by approximately 60-63% during 2D diagnostic cerebral angiography without compromising image quality, enhancing procedural safety. <h3>Disclosure of Interest</h3> no.

Authors

Institutions

Publication Details

Journal
Abstracts
Published
2024-09-01
DOI
https://doi.org/10.1136/jnis-2024-esmint.195
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

P160 Innovative reduction in radiation dose for diagnostic cerebral angiography: the efficacy of OPTIQ with IQ36

Woochul Cho, Yong Sam Shin, Jai Ho Choi, Bum Soo Kim
Abstracts
Radiation Dose and Imaging
article

P160 Innovative reduction in radiation dose for diagnostic cerebral angiography: the efficacy of OPTIQ with IQ36

Woochul Cho, Yong Sam Shin, Jai Ho Choi, Bum Soo Kim
article en

Abstract

<h3>Introduction</h3> The OPTIQ, an advanced 2D imaging pipeline, automatically adjusts exposure parameters to maintain predefined image quality during cerebral angiography while aiming to reduce radiation dose. <h3>Aim of Study</h3> This study evaluates the efficacy of OPTIQ in diminishing radiation exposure without compromising image quality in low-dose diagnostic cerebral angiography. <h3>Methods</h3> We conducted a retrospective analysis of 2D diagnostic cerebral angiography data from 100 patients (68 female, 32 male; median age 58, range 31–80 years) assessed for intracranial aneurysms. Image quality was compared between 50 patients using conventional settings and 50 using OPTIQ with an Image Quality level of 36 (IQ36), evaluated by two blinded neurointerventionists on a four-point scale across four regions. Noninferiority in image quality and reductions in radiation dose were statistically analyzed using dose-area product (DAP) and air kerma (AK) measurements. <h3>Results</h3> The OPTIQ group demonstrated noninferior image quality (Reader 1: 3.50 ± 0.38; Reader 2: 3.70 ± 0.26) compared to the conventional group (Reader 1: 3.53 ± 0.37; Reader 2: 3.78 ± 0.19). Interrater reliability was established with a Cronbach’s alpha of 0.835. DAP and AK for 2D DSA in anteroposterior and lateral views were significantly lower in the OPTIQ group (1.94 Gy-cm² and 0.012 Gy, respectively) compared to the conventional group (5.28 Gy-cm² and 0.031 Gy, respectively), marking a dose reduction of 63.26% for DAP and 60.25% for AK. <h3>Conclusion</h3> OPTIQ with IQ36 significantly reduces radiation exposure by approximately 60-63% during 2D diagnostic cerebral angiography without compromising image quality, enhancing procedural safety. <h3>Disclosure of Interest</h3> no.

Abstracts
The Catholic University of Korea Seoul St. Mary's Hospital (KR)
Openalex Percentile: Top 75%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.