P023 Cyclic aspiration in mechanical thrombectomy: influencing factors and experimental validation

<h3>Introduction</h3> Mechanical thrombectomy is a fundamental intervention for acute ischemic stroke treatment. While conventional techniques are effective, cyclic aspiration (CyA) shows potential for better recanalization rates. <h3>Aim of Study</h3> We aim to investigate factors affecting CyA and compare them with static aspiration (StA). <h3>Methods</h3> StA setup consisted of an aspiration pump connected to pressure transducer. CyA was tested with five setups: single solenoid valve with air+saline (1) or saline alone (2) as aspiration medium; two solenoid valves with air+saline (3) as aspiration medium; complete air removal and saline feeding (4); pressurized saline feeding (5). To assess the efficacy of clot ingestion, the pressure transducer was replaced with a distal aspiration catheter. Moderately stiff clot analogs (15 mm) were used to investigate the ingestion quantified as clot relative weight loss. Additionally, the aspiration flow rate was assessed for each setup. <h3>Results</h3> With CyA setup 1, the amplitude of the achieved negative pressure waves declined with increasing frequencies but progressively increased with each subsequent iteration, achieving a maximum amplitude of 81 kPa for setup 5 at 1Hz. Relative clot weight loss was significantly higher with setup 5 at 5Hz than with StA (100% vs. 37,8%; p=0.05). Aspiration flow rate was lower with CyA than with StA (setup 5 at 5Hz: 199,8ml/min vs. StA: 311ml/min; p&lt;0.01). <h3>Conclusion</h3> Cyclic aspiration with the appropriate setup may represent an encouraging innovation in mechanical thrombectomy, offering a promising pathway for improving efficacy in clot ingestion and recanalization. The observed benefits warrant confirmation in a clinical setting. <h3>Disclosure of Interest</h3> no.

Authors

Institutions

Publication Details

Journal
Abstracts
Published
2024-09-01
DOI
https://doi.org/10.1136/jnis-2024-esmint.60
Primary Topic
Acute Ischemic Stroke Management
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

P023 Cyclic aspiration in mechanical thrombectomy: influencing factors and experimental validation

Riccardo Tiberi, Alejandro Tomasello, Marc Ribó, Jiahui Li et al.
Abstracts
Acute Ischemic Stroke Management
article

P023 Cyclic aspiration in mechanical thrombectomy: influencing factors and experimental validation

Riccardo Tiberi, Alejandro Tomasello, Marc Ribó, Jiahui Li, Magda Jablonska
article en

Abstract

<h3>Introduction</h3> Mechanical thrombectomy is a fundamental intervention for acute ischemic stroke treatment. While conventional techniques are effective, cyclic aspiration (CyA) shows potential for better recanalization rates. <h3>Aim of Study</h3> We aim to investigate factors affecting CyA and compare them with static aspiration (StA). <h3>Methods</h3> StA setup consisted of an aspiration pump connected to pressure transducer. CyA was tested with five setups: single solenoid valve with air+saline (1) or saline alone (2) as aspiration medium; two solenoid valves with air+saline (3) as aspiration medium; complete air removal and saline feeding (4); pressurized saline feeding (5). To assess the efficacy of clot ingestion, the pressure transducer was replaced with a distal aspiration catheter. Moderately stiff clot analogs (15 mm) were used to investigate the ingestion quantified as clot relative weight loss. Additionally, the aspiration flow rate was assessed for each setup. <h3>Results</h3> With CyA setup 1, the amplitude of the achieved negative pressure waves declined with increasing frequencies but progressively increased with each subsequent iteration, achieving a maximum amplitude of 81 kPa for setup 5 at 1Hz. Relative clot weight loss was significantly higher with setup 5 at 5Hz than with StA (100% vs. 37,8%; p=0.05). Aspiration flow rate was lower with CyA than with StA (setup 5 at 5Hz: 199,8ml/min vs. StA: 311ml/min; p&lt;0.01). <h3>Conclusion</h3> Cyclic aspiration with the appropriate setup may represent an encouraging innovation in mechanical thrombectomy, offering a promising pathway for improving efficacy in clot ingestion and recanalization. The observed benefits warrant confirmation in a clinical setting. <h3>Disclosure of Interest</h3> no.

Abstracts
Hebron University (PS), Vall d'Hebron Institut de Recerca (ES), Vall d'Hebron Hospital Universitari (ES), Gdańsk Medical University (PL)
Zero hunger
Openalex Percentile: Top 78%
Acute Ischemic Stroke Management
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.