O30 Effective radiation dose reduction in diagnostic cerebral angiography using OPTIQ and disease-specific protocols

<h3>Introduction</h3> We have continuously tried various protocol adjustments, including the adoption of low-dose 2D and 3D rotational digital subtraction angiography (DSA), aimed at minimizing radiation exposure during cerebral angiography procedures. And recently used the OPTIQ, an advanced 2D imaging pipeline, automatically adjusts exposure parameters to maintain predefined image quality during cerebral angiography while aiming to reduce radiation dose. <h3>Aim of Study</h3> This study assesses the effectiveness of OPTIQ and customized angiographic protocols for different cerebral vascular pathologies. <h3>Methods</h3> A review was conducted on 465 diagnostic cerebral angiographies performed between August 2023 and January 2024. Angiography was done with utilization of three dose level of angiographic protocols (low-, mid-, and high-dose), customized for indication including aneurysm-diagnostic, treated aneurysm-follow-up, stenoocclusive disease, AVM/AVF. We analyzed radiation metrics such as dose-area product (DAP), air-kerma (AK), and examined imaging parameters tailored to each vascular condition. (Table 1) <h3>Results</h3> The results demonstrated significant radiation reductions across all procedures, with the lowest radiation exposures observed in aneurysm diagnostics (DAP 19.7 ± 10.1 Gy*cm², AK 0.188 ± 0.092 Gy). Although conditions like arteriovenous malformations (AVM) and arteriovenous fistulas (AVF) recorded higher metrics, they were considerably lower than historical averages for these procedures. (Table 2) <h3>Conclusion</h3> The strategic application of OPTIQ and disease-specific protocols has effectively reduced radiation doses for diagnostic cerebral angiography. <h3>Disclosure of Interest</h3> no.

Authors

Institutions

Publication Details

Journal
Abstracts
Published
2024-09-01
DOI
https://doi.org/10.1136/jnis-2024-esmint.30
Primary Topic
Cerebrovascular and Carotid Artery Diseases
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

O30 Effective radiation dose reduction in diagnostic cerebral angiography using OPTIQ and disease-specific protocols

Yong Sam Shin, Jai Ho Choi, W Y Cho, Bum Soo Kim
Abstracts
Cerebrovascular and Carotid Artery Diseases
article

O30 Effective radiation dose reduction in diagnostic cerebral angiography using OPTIQ and disease-specific protocols

Yong Sam Shin, Jai Ho Choi, W Y Cho, Bum Soo Kim
article en

Abstract

<h3>Introduction</h3> We have continuously tried various protocol adjustments, including the adoption of low-dose 2D and 3D rotational digital subtraction angiography (DSA), aimed at minimizing radiation exposure during cerebral angiography procedures. And recently used the OPTIQ, an advanced 2D imaging pipeline, automatically adjusts exposure parameters to maintain predefined image quality during cerebral angiography while aiming to reduce radiation dose. <h3>Aim of Study</h3> This study assesses the effectiveness of OPTIQ and customized angiographic protocols for different cerebral vascular pathologies. <h3>Methods</h3> A review was conducted on 465 diagnostic cerebral angiographies performed between August 2023 and January 2024. Angiography was done with utilization of three dose level of angiographic protocols (low-, mid-, and high-dose), customized for indication including aneurysm-diagnostic, treated aneurysm-follow-up, stenoocclusive disease, AVM/AVF. We analyzed radiation metrics such as dose-area product (DAP), air-kerma (AK), and examined imaging parameters tailored to each vascular condition. (Table 1) <h3>Results</h3> The results demonstrated significant radiation reductions across all procedures, with the lowest radiation exposures observed in aneurysm diagnostics (DAP 19.7 ± 10.1 Gy*cm², AK 0.188 ± 0.092 Gy). Although conditions like arteriovenous malformations (AVM) and arteriovenous fistulas (AVF) recorded higher metrics, they were considerably lower than historical averages for these procedures. (Table 2) <h3>Conclusion</h3> The strategic application of OPTIQ and disease-specific protocols has effectively reduced radiation doses for diagnostic cerebral angiography. <h3>Disclosure of Interest</h3> no.

Abstracts
The Catholic University of Korea Seoul St. Mary's Hospital (KR)
Good health and well-being
Openalex Percentile: Top 74%
Cerebrovascular and Carotid Artery Diseases
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.