P098 Preservation of radiopacity in ethylene-vinyl alcohol liquid embolization agents through manual agitation: a comprehensive multiparametric in-vitro study

<h3>Introduction</h3> The formation of a solid plug using Ethylene-Vinyl Alcohol liquid embolization agents (EVOH LEAs) is a standard technique in neurointerventionnal embolization procedure. This process requires a waiting-period during which the tantalum powder, which enhances radiopacity, may settle, resulting in decreased radiopacity. In daily practice, to prevent sedimentation, it is standard practice to agitate the syringe as the plug solidifies. <h3>Aim of Study</h3> The main objective of this study is to determine the sedimentation duration of tantalum and to identify the optimal method of manual agitation for maintaining satisfactory radiopacity. <h3>Methods</h3> EVOH LEA was prepared and assessed using optical imaging and fluoroscopic analysis of kinetic sedimentation. The quality of visibility, signal-to-noise ratio (SNR), and absorbance in 1mL syringes subjected to 8 agitation methods were assessed (fast or slow rotation either automated or manual, fast or slow horizontal shaking either automated or manual) at intervals of 3, 5, 7, and 10 minutes. Additionally, we tested these agitation methods at various time points during in-vitro embolization on a 3D-printed model simulating a distal intracranial artery. <h3>Results</h3> Early sedimentation resulting in a loss of radiopacity was noted under fluoroscopy within just 3 minutes after preparation in 1mL syringes. Despite varying time intervals, all agitation techniques successfully maintained satisfactory radiopacity in the syringes and throughout the embolization process. <h3>Conclusion</h3> Considering the swift sedimentation of EVOH LEA in a 1mL syringe, consistent agitation of EVOH LEA during the waiting time is recommended in clinical settings, with the choice of agitation method left to the practitioner’s discretion to ensure optimal radiopacity. <h3>Disclosure of Interest</h3> no.

Authors

Institutions

Publication Details

Journal
Abstracts
Published
2024-09-01
DOI
https://doi.org/10.1136/jnis-2024-esmint.134
Primary Topic
Medical Imaging and Pathology Studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

P098 Preservation of radiopacity in ethylene-vinyl alcohol liquid embolization agents through manual agitation: a comprehensive multiparametric in-vitro study

Jean‐François Hak, Thibault Agripnidis, Jan‐Patrick Stellmann
Abstracts
Medical Imaging and Pathology Studies
article

P098 Preservation of radiopacity in ethylene-vinyl alcohol liquid embolization agents through manual agitation: a comprehensive multiparametric in-vitro study

Jean‐François Hak, Thibault Agripnidis, Jan‐Patrick Stellmann
article en

Abstract

<h3>Introduction</h3> The formation of a solid plug using Ethylene-Vinyl Alcohol liquid embolization agents (EVOH LEAs) is a standard technique in neurointerventionnal embolization procedure. This process requires a waiting-period during which the tantalum powder, which enhances radiopacity, may settle, resulting in decreased radiopacity. In daily practice, to prevent sedimentation, it is standard practice to agitate the syringe as the plug solidifies. <h3>Aim of Study</h3> The main objective of this study is to determine the sedimentation duration of tantalum and to identify the optimal method of manual agitation for maintaining satisfactory radiopacity. <h3>Methods</h3> EVOH LEA was prepared and assessed using optical imaging and fluoroscopic analysis of kinetic sedimentation. The quality of visibility, signal-to-noise ratio (SNR), and absorbance in 1mL syringes subjected to 8 agitation methods were assessed (fast or slow rotation either automated or manual, fast or slow horizontal shaking either automated or manual) at intervals of 3, 5, 7, and 10 minutes. Additionally, we tested these agitation methods at various time points during in-vitro embolization on a 3D-printed model simulating a distal intracranial artery. <h3>Results</h3> Early sedimentation resulting in a loss of radiopacity was noted under fluoroscopy within just 3 minutes after preparation in 1mL syringes. Despite varying time intervals, all agitation techniques successfully maintained satisfactory radiopacity in the syringes and throughout the embolization process. <h3>Conclusion</h3> Considering the swift sedimentation of EVOH LEA in a 1mL syringe, consistent agitation of EVOH LEA during the waiting time is recommended in clinical settings, with the choice of agitation method left to the practitioner’s discretion to ensure optimal radiopacity. <h3>Disclosure of Interest</h3> no.

Abstracts
Laboratoire d’Imagerie Biomédicale (FR), Hôpital de la Timone (FR)
Zero hunger
Openalex Percentile: Top 74%
Medical Imaging and Pathology Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.