Two-Dimensional Mammography Imaging Techniques for Screening Women with Silicone Breast Implants: A Pilot Phantom Study

This study aimed to evaluate the impact of three two-dimensional (2D) mammographic acquisition techniques on image quality and radiation dose in the presence of silicone breast implants (BIs). Then, we propose and validate a new International Atomic Energy Agency (IAEA) phantom to reproduce these techniques. Images were acquired on a single Hologic Selenia Dimensions<sup>®</sup> unit. The mammography of the left breast of a single clinical case was included. Three methods of image acquisition were identified. They were based on misused, recommended, and reference settings. In the clinical case, image criteria scoring and the signal-to-noise ratio on breast tissue (SNR<sub>BT</sub>) were determined for two 2D projections and compared between the three techniques. The phantom study first compared the reference and misused settings by varying the AEC sensor position and, second, the recommended settings with a reduced current-time product (mAs) setting that was 13% lower. The signal-difference-to-noise ratio (<i>SDNR</i>) and detectability indexes at 0.1 mm (d' 0.1 mm) and 0.25 mm (d' 0.25 mm) were automatically quantified using ATIA software. Average glandular dose (AGD) values were collected for each acquisition. A statistical analysis was performed using Kruskal-Wallis and corrected Dunn tests (<i>p</i> < 0.05). The SNR<sub>BT</sub> was 2.6 times lower and the AGD was -18% lower with the reference settings compared to the recommended settings. The SNR<sub>BT</sub> values increased by +98% with the misused compared to the recommended settings. The AGD increased by +79% with the misused settings versus the recommended settings. The median values of the reference settings were 5.8 (IQR 5.7-5.9), 1.2 (IQR 0.0), 7.0 (IQR 6.8-7.2) and 1.2 (IQR 0.0) mGy and were significantly lower than those of the misused settings (<i>p</i> < 0.03): 7.9 (IQR 6.1-9.7), 1.6 (IQR 1.3-1.9), 9.2 (IQR 7.5-10.9) and 2.2 (IQR 1.4-3.0) mGy for the <i>SDNR</i>, d' 0.1 mm, d' 0.25 mm and the AGD, respectively. A comparison of the recommended and reduced settings showed a reduction of -6.1 ± 0.6% (<i>p</i> = 0.83), -7.7 ± 0.0% (<i>p</i> = 0.18), -6.4 ± 0.6% (<i>p</i> = 0.19) and -13.3 ± 1.1% (<i>p</i> = 0.53) for the <i>SDNR</i>, d' 0.1 mm, d' 0.25 mm and the AGD, respectively. This study showed that the IAEA phantom could be used to reproduce the three techniques for acquiring 2D mammography images in the presence of breast implants for raising awareness and for educational purposes. It could also be used to evaluate and optimize the manufacturer's recommended settings.

Authors

Institutions

Publication Details

Journal
Bioengineering
Published
2024-08-31
DOI
https://doi.org/10.3390/bioengineering11090884
Citations
4
Primary Topic
Breast Implant and Reconstruction
Type
article
Field-Weighted Citation Impact
2.29

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Two-Dimensional Mammography Imaging Techniques for Screening Women with Silicone Breast Implants: A Pilot Phantom Study

Françoise Malchair, Virginia Tsapaki, J. Zerbib, Isabelle Fitton et al.
4 citations
Bioengineering
Breast Implant and Reconstruction
2.29
article

Two-Dimensional Mammography Imaging Techniques for Screening Women with Silicone Breast Implants: A Pilot Phantom Study

Françoise Malchair, Virginia Tsapaki, J. Zerbib, Isabelle Fitton, Claire Van Ngoc Ty, Amit Kumar, Antoine Decoux, Laure Fournier, Aude Stembert
article en
4 citations

Abstract

This study aimed to evaluate the impact of three two-dimensional (2D) mammographic acquisition techniques on image quality and radiation dose in the presence of silicone breast implants (BIs). Then, we propose and validate a new International Atomic Energy Agency (IAEA) phantom to reproduce these techniques. Images were acquired on a single Hologic Selenia Dimensions<sup>®</sup> unit. The mammography of the left breast of a single clinical case was included. Three methods of image acquisition were identified. They were based on misused, recommended, and reference settings. In the clinical case, image criteria scoring and the signal-to-noise ratio on breast tissue (SNR<sub>BT</sub>) were determined for two 2D projections and compared between the three techniques. The phantom study first compared the reference and misused settings by varying the AEC sensor position and, second, the recommended settings with a reduced current-time product (mAs) setting that was 13% lower. The signal-difference-to-noise ratio (<i>SDNR</i>) and detectability indexes at 0.1 mm (d' 0.1 mm) and 0.25 mm (d' 0.25 mm) were automatically quantified using ATIA software. Average glandular dose (AGD) values were collected for each acquisition. A statistical analysis was performed using Kruskal-Wallis and corrected Dunn tests (<i>p</i> < 0.05). The SNR<sub>BT</sub> was 2.6 times lower and the AGD was -18% lower with the reference settings compared to the recommended settings. The SNR<sub>BT</sub> values increased by +98% with the misused compared to the recommended settings. The AGD increased by +79% with the misused settings versus the recommended settings. The median values of the reference settings were 5.8 (IQR 5.7-5.9), 1.2 (IQR 0.0), 7.0 (IQR 6.8-7.2) and 1.2 (IQR 0.0) mGy and were significantly lower than those of the misused settings (<i>p</i> < 0.03): 7.9 (IQR 6.1-9.7), 1.6 (IQR 1.3-1.9), 9.2 (IQR 7.5-10.9) and 2.2 (IQR 1.4-3.0) mGy for the <i>SDNR</i>, d' 0.1 mm, d' 0.25 mm and the AGD, respectively. A comparison of the recommended and reduced settings showed a reduction of -6.1 ± 0.6% (<i>p</i> = 0.83), -7.7 ± 0.0% (<i>p</i> = 0.18), -6.4 ± 0.6% (<i>p</i> = 0.19) and -13.3 ± 1.1% (<i>p</i> = 0.53) for the <i>SDNR</i>, d' 0.1 mm, d' 0.25 mm and the AGD, respectively. This study showed that the IAEA phantom could be used to reproduce the three techniques for acquiring 2D mammography images in the presence of breast implants for raising awareness and for educational purposes. It could also be used to evaluate and optimize the manufacturer's recommended settings.

BioengineeringVol. 11(9)
International Atomic Energy Agency (AT), Inserm (FR), Université Paris Cité (FR), Assistance Publique – Hôpitaux de Paris (FR), Hôpital Européen Georges-Pompidou (FR), Hôpital Européen (FR), Paris Cardiovascular Research Center (FR)
International Atomic Energy Agency
Openalex Percentile: Top 12%
Breast Implant and Reconstruction
2.29
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.