Registration accuracy comparing different rendering techniques on local vs external virtual 3D liver model reconstruction for vascular landmark setting by intraoperative ultrasound in augmented reality navigated liver resection
Despite basing on different rendering techniques both local and external vir3DLivers have comparable registration accuracy, while LPV-based registrations significantly outperform RPV-based ones in accuracy. Higher accuracy can be assumed within distances of up to a few centimeters around the ROI.
Authors
- Roger Wahba (ORCID: https://orcid.org/0000-0003-0037-8620)
- Alexander C. Bunck (ORCID: https://orcid.org/0000-0003-0986-0042)
- Martin Hellmich (ORCID: https://orcid.org/0000-0001-5174-928X)
- Michael Nikolaus Thomas (ORCID: https://orcid.org/0000-0002-4121-3091)
- Christiane J. Bruns (ORCID: https://orcid.org/0000-0001-6590-8181)
- Dirk L. Stippel (ORCID: https://orcid.org/0000-0002-1107-0907)
- Matthias Peterhans
- Rabi Raj Datta (ORCID: https://orcid.org/0000-0002-6829-7379)
- Nonkoh J. Sheriff
Institutions
- University of Cologne (DE)
- University Hospital Cologne (DE)
- Helios Hospital Berlin-Buch (DE)
Publication Details
- Journal
- Langenbeck s Archives of Surgery
- Published
- 2024-09-03
- DOI
- https://doi.org/10.1007/s00423-024-03456-z
- Citations
- 3
- Primary Topic
- Surgical Simulation and Training
- Type
- article
- Field-Weighted Citation Impact
- 1.72
Funders
- Universitätsklinikum Köln