Evaluation and Comparison of the Dose Received by the Mandible, Maxilla, and Teeth in Two Methods of Three-dimensional Conformal Radiation Therapy and Helical Tomotherapy
The findings of the comparison of the two treatment approaches revealed that the HT method was more than 50% more effective than the 3D-CRT method in sustaining organs at risk (OARs) and the target volume dose. In general, dosimetric coverage, homogeneity, conformity indices, and the absence of cold and hot patches showed that HT produced targets with greater accuracy than 3D-CRT. In addition, HT outperformed 3D-CRT in protecting important structures (OARs). HT as a result has the potential to be a more effective method of treatment for those with H and NC and involvement of regional lymph nodes.
Authors
- Mohsen Saeb (ORCID: https://orcid.org/0000-0002-6522-3497)
- Pegah Moradi Khaniabadi (ORCID: https://orcid.org/0000-0001-7245-1087)
- Bita Moradi Khaniabadi (ORCID: https://orcid.org/0000-0003-1185-838X)
- Nadia Najafizade
- Daryoush Shahbazi‐Gahrouei (ORCID: https://orcid.org/0000-0002-8538-1898)
- Zahra Pourparvar
Institutions
- Isfahan University of Medical Sciences (IR)
- Sultan Qaboos University (OM)
Publication Details
- Journal
- Journal of Medical Signals & Sensors
- Published
- 2024-09-01
- DOI
- https://doi.org/10.4103/jmss.jmss_42_23
- Citations
- 1
- Primary Topic
- Head and Neck Cancer Studies
- Type
- article
- Field-Weighted Citation Impact
- 0.58
Funders
- Isfahan University of Medical Sciences