How Does the Stress in the Fixation Device Change during Different Stages of Bone Healing in the Treatment of Fractures? A Finite Element Study of External Fixation for Tibial Fractures
During the bone healing process, stress is transferred between the external fixation frame and the bone. As bone healing advances, the stress on the connecting rods and screws of the external fixation frame decreases, and the amplitude of stress changes diminishes. When complete and robust fusion is achieved, stress variations stabilize, and the location of maximum stress on the external fixation frame remains unchanged. The intersections of the lowest screw and the bone cortex, as well as the second screw and the connecting rod, can serve as sensitive points for monitoring the degree of bone healing.
Authors
- Yi Deng (ORCID: https://orcid.org/0000-0002-1765-5244)
- Xuehai Jia (ORCID: https://orcid.org/0000-0002-0341-196X)
- Bin Luo (ORCID: https://orcid.org/0000-0002-8721-4460)
- Jun Wen (ORCID: https://orcid.org/0000-0003-2929-1270)
- Litai Ma
- Kerui Zhang
- Yi Yang (ORCID: https://orcid.org/0009-0007-6082-3307)
- Changyong Shen
Institutions
- Southwest University of Science and Technology (CN)
- Sichuan University (CN)
- West China Hospital of Sichuan University (CN)
Publication Details
- Journal
- Orthopaedic Surgery
- Published
- 2024-09-02
- DOI
- https://doi.org/10.1111/os.14195
- Citations
- 2
- Primary Topic
- Bone fractures and treatments
- Type
- article
- Field-Weighted Citation Impact
- 0.77
Funders
- Chengdu Science and Technology Program