A comparison of white-to-white measurements obtained by Anterion AS-OCT versus three optical devices in refractive surgery candidates

Background and ObjectivesTo assess the agreement between Anterion AS-OCT and three optical devices in measuring the white-to-white (WTW) diameter in candidates for refractive surgery.MethodsIn this cross-sectional prospective study, 129 right eyes of 129 normal individuals underwent consecutive scans with the Anterion, the Pentacam AXL, the IOLMaster 700, and the Orbscan IIz. Mean difference (MD), 95% limits of agreement (LoA), and concordance correlation coefficient (CCC) were calculated to assess agreement and interchangeability.ResultsThe mean age of participants was 30.4 ± 5.9 (range: 21-47) years. The mean WTW distance measured by the different instruments was as follows: 12.00 ± 0.42 mm (range, 10.57 to 13.05) using the Anterion, 11.87 ± 0.34 (11.10 to 12.80) using the Pentacam, 12.12 ± 0.44 (11.00 to 13.30) using the IOLMaster, and 11.73 ± 0.37 (11.00 to 13.10) using the Orbscan. The MD and 95% LoA for Anterion vs. Pentacam, IOLMaster, or Orbscan were [0.11; -0.31 to 0.54 mm], [-0.13; -0.93 to 0.66], and [0.25; -0.28 to 0.78], respectively. The corresponding CCCs were 0.803, 0.514, and 0.631.ConclusionsThis study found weak agreement between Anterion and Pentacam AXL, IOLMaster 700, and Orbscan IIz devices regarding WTW distance measurements in refractive surgery candidates. Therefore, it is not recommended to use Anterion's WTW measurements interchangeably with the other three devices, particularly for phakic intraocular lens sizing.

Authors

Institutions

Publication Details

Journal
European Journal of Ophthalmology
Published
2024-09-02
DOI
https://doi.org/10.1177/11206721241281876
Citations
2
Primary Topic
Corneal surgery and disorders
Type
article
Field-Weighted Citation Impact
0.81
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A comparison of white-to-white measurements obtained by Anterion AS-OCT versus three optical devices in refractive surgery candidates

Kia Salouti, Maryam Ghoreyshi, M. Hossein Nowroozzadeh, Ramin Salouti et al.
2 citations
European Journal of Ophthalmology
Corneal surgery and disorders
0.81
article

A comparison of white-to-white measurements obtained by Anterion AS-OCT versus three optical devices in refractive surgery candidates

Kia Salouti, Maryam Ghoreyshi, M. Hossein Nowroozzadeh, Ramin Salouti, Hesam Hashemian, Masoud Khorrami‐Nejad, Ali Azizi, Qassim Alkhafaji
article en
2 citations

Abstract

Background and ObjectivesTo assess the agreement between Anterion AS-OCT and three optical devices in measuring the white-to-white (WTW) diameter in candidates for refractive surgery.MethodsIn this cross-sectional prospective study, 129 right eyes of 129 normal individuals underwent consecutive scans with the Anterion, the Pentacam AXL, the IOLMaster 700, and the Orbscan IIz. Mean difference (MD), 95% limits of agreement (LoA), and concordance correlation coefficient (CCC) were calculated to assess agreement and interchangeability.ResultsThe mean age of participants was 30.4 ± 5.9 (range: 21-47) years. The mean WTW distance measured by the different instruments was as follows: 12.00 ± 0.42 mm (range, 10.57 to 13.05) using the Anterion, 11.87 ± 0.34 (11.10 to 12.80) using the Pentacam, 12.12 ± 0.44 (11.00 to 13.30) using the IOLMaster, and 11.73 ± 0.37 (11.00 to 13.10) using the Orbscan. The MD and 95% LoA for Anterion vs. Pentacam, IOLMaster, or Orbscan were [0.11; -0.31 to 0.54 mm], [-0.13; -0.93 to 0.66], and [0.25; -0.28 to 0.78], respectively. The corresponding CCCs were 0.803, 0.514, and 0.631.ConclusionsThis study found weak agreement between Anterion and Pentacam AXL, IOLMaster 700, and Orbscan IIz devices regarding WTW distance measurements in refractive surgery candidates. Therefore, it is not recommended to use Anterion's WTW measurements interchangeably with the other three devices, particularly for phakic intraocular lens sizing.

European Journal of OphthalmologyVol. 35(3)
University of British Columbia (CA), Shiraz University of Medical Sciences (IR), Farabi Hospital (IR), Tehran University of Medical Sciences (IR)
Openalex Percentile: Top 26%
Corneal surgery and disorders
0.81
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.