Calibration verification of an underwater hyperspectral imaging push broom instrument to measure light in absolute units and field demonstration

The push broom design of an underwater hyperspectral imaging (UHI) instrument makes it possible to measure angle-resolved spectral radiance L ( λ ) in a plane. We describe the characterization of a commercial UHI instrument (UHI-4, Ecotone AS, Norway) and the spectral, geometric, and radiometric calibration transfer for measuring L ( λ ) in absolute units [µWcm −2 nm −1 sr −1 ]. We present a low-cost instrument characterization approach that is intended to be easily replicated for other users to perform their own calibration transfer. Cross-calibration with a RAMSES-ARC spectroradiometer (TriOS, Germany) in air and in water shows good linear correlation across the observed spectral range and over four orders of magnitude in dynamic range, with the UHI instrument providing higher sensitivity overall.

Authors

Institutions

Publication Details

Journal
Applied Optics
Published
2024-09-03
DOI
https://doi.org/10.1364/ao.532003
Primary Topic
bioluminescence and chemiluminescence research
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Calibration verification of an underwater hyperspectral imaging push broom instrument to measure light in absolute units and field demonstration

Geir Johnsen, Jørgen Berge, David McKee, Bernhard Schartmüller
Applied Optics
bioluminescence and chemiluminescence research
article

Calibration verification of an underwater hyperspectral imaging push broom instrument to measure light in absolute units and field demonstration

Geir Johnsen, Jørgen Berge, David McKee, Bernhard Schartmüller
article en

Abstract

The push broom design of an underwater hyperspectral imaging (UHI) instrument makes it possible to measure angle-resolved spectral radiance L ( λ ) in a plane. We describe the characterization of a commercial UHI instrument (UHI-4, Ecotone AS, Norway) and the spectral, geometric, and radiometric calibration transfer for measuring L ( λ ) in absolute units [µWcm −2 nm −1 sr −1 ]. We present a low-cost instrument characterization approach that is intended to be easily replicated for other users to perform their own calibration transfer. Cross-calibration with a RAMSES-ARC spectroradiometer (TriOS, Germany) in air and in water shows good linear correlation across the observed spectral range and over four orders of magnitude in dynamic range, with the UHI instrument providing higher sensitivity overall.

Applied OpticsVol. 63(27)
University of Strathclyde (GB), Norwegian University of Science and Technology (NO), University Centre in Svalbard (SJ)
Sight Research UK, Science Foundation Ireland, Norges Forskningsråd, Natural Environment Research Council
Life below water
Openalex Percentile: Top 87%
bioluminescence and chemiluminescence research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.