Optimization of motion-corrected liver diffusion-weighted imaging at 3 Tesla (3T): incorporating complex averaging and reparametrized sinc fatsat pulse with optimized water excitation pulse

Background: In liver diffusion-weighted imaging (DWI), single-shot echo-planar imaging (SS-EPI) sequences are susceptible to motion artifacts, resulting in image blurring and decreased lesion detection rates. This study aimed to develop and optimize a motion-corrected (MOCO) technique for liver DWI at 3 Tesla (3T). The technique incorporates motion correction, complex averaging, and a combination of a reparametrized sinc fatsat pulse with an optimized water excitation pulse. Methods: This prospective cross-sectional study performed at Fujian Medical University Union Hospital included 42 healthy volunteers who underwent four SS-EPI DWI sequences on a 3T magnetic resonance imaging (MRI) system between January 2023 and March 2023. The sequences included a navigator-triggered (NT) MOCO-DWI, two free-breathing (FB) MOCO-DWI, and an FB conventional DWI (FB cDWI) sequence. Motion correction and complex averaging were performed for both MOCO-DWI sequences, and fat suppression was achieved using either a sinc fatsat pulse with optimized water excitation or a conventional spectral attenuated inversion recovery (SPAIR) pulse. Liver signal-to-noise ratio (SNR) was measured at b=1,000 s/mm2. Qualitative parameters were independently evaluated by three radiologists using 5-point Likert scales. Quantitative parameters were assessed using the Kolmogorov-Smirnov test, and variance homogeneity was assessed using Levene’s test. Regarding the qualitative analysis, the Friedman test was used to compare subjective scores among the four techniques. Results: The SNRs of the liver were significantly higher with FB MOCO-DWI compared to the other EPI DWI sequences at b=1,000 s/mm2 (P<0.05). In the superior-inferior direction, the SNRs of the inferior level of the liver were higher than those of the superior level in NT MOCO-DWI. The qualitative results showed significantly higher ratings for NT MOCO-DWI and FB MOCO-DWI compared to the other EPI DWI sequences at b=1,000 s/mm2 (P<0.05). Regarding the apparent diffusion coefficient (ADC) quantification, the ADC values of the left lobe were higher than those of the right lobe in all four techniques. Conclusions: The proposed EPI DWI technique, incorporating motion correction, complex averaging, and a modified fat suppression scheme using spectral fat saturation and binomial water excitation, was found to be clinically feasible for liver MRI. The FB MOCO-DWI sequence, with its superior SNR and excellent image quality, is recommended for liver DW imaging at 3T in clinical routine.

Authors

Institutions

Publication Details

Journal
Quantitative Imaging in Medicine and Surgery
Published
2024-09-01
DOI
https://doi.org/10.21037/qims-24-340
Citations
1
Primary Topic
MRI in cancer diagnosis
Type
article
Field-Weighted Citation Impact
0.40
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Optimization of motion-corrected liver diffusion-weighted imaging at 3 Tesla (3T): incorporating complex averaging and reparametrized sinc fatsat pulse with optimized water excitation pulse

Enshuang Zheng, Thomas Benkert, Guijin Li, Zhangli Xing et al.
1 citations
Quantitative Imaging in Medicine and Surgery
MRI in cancer diagnosis
0.40
article

Optimization of motion-corrected liver diffusion-weighted imaging at 3 Tesla (3T): incorporating complex averaging and reparametrized sinc fatsat pulse with optimized water excitation pulse

Enshuang Zheng, Thomas Benkert, Guijin Li, Zhangli Xing, Zhiyong Chen, Caixia Fu, Bin Sun, Mingcong Luo, Yunjing Xue
article en
1 citations

Abstract

Background: In liver diffusion-weighted imaging (DWI), single-shot echo-planar imaging (SS-EPI) sequences are susceptible to motion artifacts, resulting in image blurring and decreased lesion detection rates. This study aimed to develop and optimize a motion-corrected (MOCO) technique for liver DWI at 3 Tesla (3T). The technique incorporates motion correction, complex averaging, and a combination of a reparametrized sinc fatsat pulse with an optimized water excitation pulse. Methods: This prospective cross-sectional study performed at Fujian Medical University Union Hospital included 42 healthy volunteers who underwent four SS-EPI DWI sequences on a 3T magnetic resonance imaging (MRI) system between January 2023 and March 2023. The sequences included a navigator-triggered (NT) MOCO-DWI, two free-breathing (FB) MOCO-DWI, and an FB conventional DWI (FB cDWI) sequence. Motion correction and complex averaging were performed for both MOCO-DWI sequences, and fat suppression was achieved using either a sinc fatsat pulse with optimized water excitation or a conventional spectral attenuated inversion recovery (SPAIR) pulse. Liver signal-to-noise ratio (SNR) was measured at b=1,000 s/mm2. Qualitative parameters were independently evaluated by three radiologists using 5-point Likert scales. Quantitative parameters were assessed using the Kolmogorov-Smirnov test, and variance homogeneity was assessed using Levene’s test. Regarding the qualitative analysis, the Friedman test was used to compare subjective scores among the four techniques. Results: The SNRs of the liver were significantly higher with FB MOCO-DWI compared to the other EPI DWI sequences at b=1,000 s/mm2 (P<0.05). In the superior-inferior direction, the SNRs of the inferior level of the liver were higher than those of the superior level in NT MOCO-DWI. The qualitative results showed significantly higher ratings for NT MOCO-DWI and FB MOCO-DWI compared to the other EPI DWI sequences at b=1,000 s/mm2 (P<0.05). Regarding the apparent diffusion coefficient (ADC) quantification, the ADC values of the left lobe were higher than those of the right lobe in all four techniques. Conclusions: The proposed EPI DWI technique, incorporating motion correction, complex averaging, and a modified fat suppression scheme using spectral fat saturation and binomial water excitation, was found to be clinically feasible for liver MRI. The FB MOCO-DWI sequence, with its superior SNR and excellent image quality, is recommended for liver DW imaging at 3T in clinical routine.

Quantitative Imaging in Medicine and SurgeryVol. 14(9)
Fujian Medical University (CN), Siemens Healthcare (United States) (US), Siemens Healthineers (Germany) (DE), Union Hospital (CN), Siemens (China) (CN)
Clean water and sanitation
Openalex Percentile: Top 35%
MRI in cancer diagnosis
0.40
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.