Superior Measurement Accuracy of Digital Thickness Gauge Versus Digital Vernier Caliper in Determining Venous Tissue Thickness

Background In determining mechanical characteristics, the accuracy of the thickness of the specimens can influence the biomechanical behavior, especially in the case of human tissues, where there is an important variability. This study aims to compare the accuracy of two routine measuring instruments, i.e., the digital vernier caliper and the digital thickness gauge, when measuring the thickness of venous specimens multiple times. Methodology In this study, we used 12 tubular vena cava specimens obtained from common breed pigs aged 18-24 weeks at the time of sacrifice from a local slaughterhouse. The measurements were performed using a digital vernier caliper (Multicomp PRO MP012475) for the first four protocols and a digital thickness gauge (Mitutoyo 547-500S) for the fifth protocol. In the first protocol, three measurements were taken on the same side, and their average was recorded as the sample thickness. The second protocol involved taking measurements on two opposite sides, and the average of these measurements was recorded as the sample thickness. In the third protocol, the thickness of each side was measured at its midpoint, and the average of the four measurements was recorded as the sample thickness. In the last protocol using a digital vernier caliper, the thickness of the vernier specimens was calculated as the average of the measurements taken at each corner of the square sample. Finally, for the fifth protocol, three consecutive measurements were taken using the digital thickness gauge, and their average was recorded as the final thickness of the sample. Results In the first protocol, we observed lower values during the first measurement in comparison to the second (0.409 ± 0.063 vs. 0.536 ± 0.064, p < 0.0001) and the third (0.409 ± 0.063 vs. 0.528 ± 0.055, p = 0.0001). Moreover, with the second protocol, we observed lower values during the first two measurements in comparison to the third measurement (p = 0.0279 and p = 0.0054). Regarding protocols three and four, we recorded higher values for the second and third measurements than the first one, with higher values for the third measurement than the second one. In the fifth protocol, there were no significant statistical differences between the three consecutive measurements (p = 0.953, p = 0.742, and p = 0.897). Further, we examined the variations in sample thickness determined using each of the protocols proposed for the digital vernier caliper, as well as the values obtained with the digital thickness gauge protocol. As a result, during the first and second measurements, we observed lower thickness values for the venous wall samples using the first four protocols compared to the fifth protocol (for all p < 0.05). However, no differences were noted between the five protocols during the third measurement. Conclusions The digital thickness gauge Mitutoyo 547-500S provided superior accuracy with no difference between three successive measurements of venous wall thickness, regardless of the examiner's experience. Accurately determining the thickness of venous specimens is crucial for calculating the tissue's biomechanical properties.

Authors

Institutions

Publication Details

Journal
Cureus
Published
2024-09-02
DOI
https://doi.org/10.7759/cureus.68442
Primary Topic
Venous Thromboembolism Diagnosis and Management
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Superior Measurement Accuracy of Digital Thickness Gauge Versus Digital Vernier Caliper in Determining Venous Tissue Thickness

Traian V. Chirilă, Claudiu Constantin Ciucanu, Réka Bartus, Carmen Corina Radu et al.
Cureus
Venous Thromboembolism Diagnosis and Management
article

Superior Measurement Accuracy of Digital Thickness Gauge Versus Digital Vernier Caliper in Determining Venous Tissue Thickness

Traian V. Chirilă, Claudiu Constantin Ciucanu, Réka Bartus, Carmen Corina Radu, Eliza Mihaela Arbănași, Eliza Russu, Emil Marian Arbănași, Alexandru Petru Ion, Adrian Vasile Mureșan, Alexandra Asztalos
article en

Abstract

Background In determining mechanical characteristics, the accuracy of the thickness of the specimens can influence the biomechanical behavior, especially in the case of human tissues, where there is an important variability. This study aims to compare the accuracy of two routine measuring instruments, i.e., the digital vernier caliper and the digital thickness gauge, when measuring the thickness of venous specimens multiple times. Methodology In this study, we used 12 tubular vena cava specimens obtained from common breed pigs aged 18-24 weeks at the time of sacrifice from a local slaughterhouse. The measurements were performed using a digital vernier caliper (Multicomp PRO MP012475) for the first four protocols and a digital thickness gauge (Mitutoyo 547-500S) for the fifth protocol. In the first protocol, three measurements were taken on the same side, and their average was recorded as the sample thickness. The second protocol involved taking measurements on two opposite sides, and the average of these measurements was recorded as the sample thickness. In the third protocol, the thickness of each side was measured at its midpoint, and the average of the four measurements was recorded as the sample thickness. In the last protocol using a digital vernier caliper, the thickness of the vernier specimens was calculated as the average of the measurements taken at each corner of the square sample. Finally, for the fifth protocol, three consecutive measurements were taken using the digital thickness gauge, and their average was recorded as the final thickness of the sample. Results In the first protocol, we observed lower values during the first measurement in comparison to the second (0.409 ± 0.063 vs. 0.536 ± 0.064, p < 0.0001) and the third (0.409 ± 0.063 vs. 0.528 ± 0.055, p = 0.0001). Moreover, with the second protocol, we observed lower values during the first two measurements in comparison to the third measurement (p = 0.0279 and p = 0.0054). Regarding protocols three and four, we recorded higher values for the second and third measurements than the first one, with higher values for the third measurement than the second one. In the fifth protocol, there were no significant statistical differences between the three consecutive measurements (p = 0.953, p = 0.742, and p = 0.897). Further, we examined the variations in sample thickness determined using each of the protocols proposed for the digital vernier caliper, as well as the values obtained with the digital thickness gauge protocol. As a result, during the first and second measurements, we observed lower thickness values for the venous wall samples using the first four protocols compared to the fifth protocol (for all p < 0.05). However, no differences were noted between the five protocols during the third measurement. Conclusions The digital thickness gauge Mitutoyo 547-500S provided superior accuracy with no difference between three successive measurements of venous wall thickness, regardless of the examiner's experience. Accurately determining the thickness of venous specimens is crucial for calculating the tissue's biomechanical properties.

CureusVol. 16(9)
Queensland University of Technology (AU), The University of Queensland (AU), Universitatea de Medicină, Farmacie, Științe și Tehnologie „George Emil Palade” din Târgu Mureș (RO), Queensland Eye Institute (AU), Spitalul Clinic Judetean de Urgenta Târgu Mureş (RO), Spitalul Clinic Judetean Mures (RO)
Openalex Percentile: Top 71%
Venous Thromboembolism Diagnosis and Management
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.