Validity and Reliability of Movesense HR+ ECG Measurements for High-Intensity Running and Cycling

Low-cost, portable devices capable of accurate physiological measurements are attractive tools for coaches, athletes, and practitioners. The purpose of this study was primarily to establish the validity and reliability of Movesense HR+ ECG measurements compared to the criterion three-lead ECG, and secondarily, to test the industry leader Garmin HRM. Twenty-one healthy adults participated in running and cycling incremental test protocols to exhaustion, both with rest before and after. Movesense HR+ demonstrated consistent and accurate R-peak detection, with an overall sensitivity of 99.7% and precision of 99.6% compared to the criterion; Garmin HRM sensitivity and precision were 84.7% and 87.7%, respectively. Bland-Altman analysis compared to the criterion indicated mean differences (SD) in RR' intervals of 0.23 (22.3) ms for Movesense HR+ at rest and 0.38 (18.7) ms during the incremental test. The mean difference for Garmin HRM-Pro at rest was -8.5 (111.5) ms and 27.7 (128.7) ms for the incremental test. The incremental test correlation was very strong (<i>r</i> = 0.98) between Movesense HR+ and criterion, and moderate (<i>r</i> = 0.66) for Garmin HRM-Pro. This study developed a robust peak detection algorithm and data collection protocol for Movesense HR+ and established its validity and reliability for ECG measurement.

Authors

Institutions

Publication Details

Journal
Sensors
Published
2024-09-02
DOI
https://doi.org/10.3390/s24175713
Citations
7
Primary Topic
Cardiovascular Effects of Exercise
Type
article
Field-Weighted Citation Impact
2.81
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Validity and Reliability of Movesense HR+ ECG Measurements for High-Intensity Running and Cycling

Peter Lamb, Enzo Allevard, Haye Kamstra, Raúl Martín Gómez et al.
7 citations
Sensors
Cardiovascular Effects of Exercise
2.81
article

Validity and Reliability of Movesense HR+ ECG Measurements for High-Intensity Running and Cycling

Peter Lamb, Enzo Allevard, Haye Kamstra, Raúl Martín Gómez, James David Cotter
article en
7 citations

Abstract

Low-cost, portable devices capable of accurate physiological measurements are attractive tools for coaches, athletes, and practitioners. The purpose of this study was primarily to establish the validity and reliability of Movesense HR+ ECG measurements compared to the criterion three-lead ECG, and secondarily, to test the industry leader Garmin HRM. Twenty-one healthy adults participated in running and cycling incremental test protocols to exhaustion, both with rest before and after. Movesense HR+ demonstrated consistent and accurate R-peak detection, with an overall sensitivity of 99.7% and precision of 99.6% compared to the criterion; Garmin HRM sensitivity and precision were 84.7% and 87.7%, respectively. Bland-Altman analysis compared to the criterion indicated mean differences (SD) in RR' intervals of 0.23 (22.3) ms for Movesense HR+ at rest and 0.38 (18.7) ms during the incremental test. The mean difference for Garmin HRM-Pro at rest was -8.5 (111.5) ms and 27.7 (128.7) ms for the incremental test. The incremental test correlation was very strong (<i>r</i> = 0.98) between Movesense HR+ and criterion, and moderate (<i>r</i> = 0.66) for Garmin HRM-Pro. This study developed a robust peak detection algorithm and data collection protocol for Movesense HR+ and established its validity and reliability for ECG measurement.

SensorsVol. 24(17)
University of Auckland (NZ), University of Otago (NZ), Vrije Universiteit Amsterdam (NL)
Industry, innovation and infrastructure
Openalex Percentile: Top 8%
Cardiovascular Effects of Exercise
2.81
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Validity and Reliability of Movesense HR+ ECG Measurements for High-Intensity Running and Cycling — Peter Lamb, Enzo Allevard, et al. · Sensors (2024) | TGRS Research Map | TGRS