Formaldehyde is a major indoor air pollutant. Accurate analysis of airborne formaldehyde is essential for effective monitoring and risk assessment. For many formaldehyde analysis techniques, the accuracy of the results depends on the calibration system. Existing formaldehyde calibration methods often face limitations, such as instability over time or limited concentration range. They also lack practicality for on-field calibration either because of bulkiness or because of high gas consumption. To address these challenges, a device using a paraformaldehyde permeation tube within a custom-built, temperature-controlled system was designed and evaluated. The device weighs less than 3 kg and operates using a nitrogen source. Its low gas consumption (30–100 mL min−1) enables the use of a portable gas cylinder of pure nitrogen (<2 kg) as gas supply. Formaldehyde concentrations ranging from 8.3 to 464 µg/m−3(−|-) were successfully generated, demonstrating the device’s versatility. A custom-built high-emission permeation tube could generate concentrations from 5 940 to 49 471 µg/m−3(−|-) using the same system. Additionally, the emission rate is independent of the flow rate and follows Antoine’s law with respect to temperature, enabling accurate prediction of concentration in various conditions over the evaluated range and beyond. Furthermore, the study confirms the long-term stability (1 year) of the low-emission paraformaldehyde permeation tube, with emission rates remaining mostly within ± 5 % of the average value for each temperature. The temperature-controlled permeation system approach enabled rapid stabilisation of formaldehyde generation at various concentrations and low flow rates, eliminating the need for additional dilution equipment and facilitating on-site instrument calibration at various gas concentrations. This reliable and efficient portable formaldehyde generator offers a valuable solution for field calibration, providing accurate and stable reference gas for diverse analytical needs.

Authors

Institutions

Publication Details

Journal
Microchemical Journal
Published
2024-09-02
DOI
https://doi.org/10.1016/j.microc.2024.111544
Citations
3
Primary Topic
Indoor Air Quality and Microbial Exposure
Type
article
Field-Weighted Citation Impact
0.94

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Portable temperature-controlled permeation device for the generation of formaldehyde gas standard

Audrey Grandjean, Anaïs Becker, Michel Wolf, Jean-Philippe Amiet et al.
3 citations
Microchemical Journal
Indoor Air Quality and Microbial Exposure
0.94
article

Portable temperature-controlled permeation device for the generation of formaldehyde gas standard

Audrey Grandjean, Anaïs Becker, Michel Wolf, Jean-Philippe Amiet, Stéphane Le Calvé, Christophe Sutter, Damien Bazin, Coralie Kustner, Romain Severac
article en
3 citations

Abstract

Formaldehyde is a major indoor air pollutant. Accurate analysis of airborne formaldehyde is essential for effective monitoring and risk assessment. For many formaldehyde analysis techniques, the accuracy of the results depends on the calibration system. Existing formaldehyde calibration methods often face limitations, such as instability over time or limited concentration range. They also lack practicality for on-field calibration either because of bulkiness or because of high gas consumption. To address these challenges, a device using a paraformaldehyde permeation tube within a custom-built, temperature-controlled system was designed and evaluated. The device weighs less than 3 kg and operates using a nitrogen source. Its low gas consumption (30–100 mL min−1) enables the use of a portable gas cylinder of pure nitrogen (<2 kg) as gas supply. Formaldehyde concentrations ranging from 8.3 to 464 µg/m−3(−|-) were successfully generated, demonstrating the device’s versatility. A custom-built high-emission permeation tube could generate concentrations from 5 940 to 49 471 µg/m−3(−|-) using the same system. Additionally, the emission rate is independent of the flow rate and follows Antoine’s law with respect to temperature, enabling accurate prediction of concentration in various conditions over the evaluated range and beyond. Furthermore, the study confirms the long-term stability (1 year) of the low-emission paraformaldehyde permeation tube, with emission rates remaining mostly within ± 5 % of the average value for each temperature. The temperature-controlled permeation system approach enabled rapid stabilisation of formaldehyde generation at various concentrations and low flow rates, eliminating the need for additional dilution equipment and facilitating on-site instrument calibration at various gas concentrations. This reliable and efficient portable formaldehyde generator offers a valuable solution for field calibration, providing accurate and stable reference gas for diverse analytical needs.

Microchemical JournalVol. 206
Centre National de la Recherche Scientifique (FR), Institut de Chimie et Procédés pour l'Energie, l'Environnement et la Santé (FR), Université de Strasbourg (FR)
Association Nationale de la Recherche et de la Technologie
Clean water and sanitation
Openalex Percentile: Top 25%
Indoor Air Quality and Microbial Exposure
0.94
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.