Exploring aortic morphology and determining variable-distance insertion lengths for fluoroscopy-free resuscitative endovascular balloon occlusion of the aorta (REBOA)
This is the first study to compare aortic morphology and intravascular distances between combined sex and age-based subgroups. As zone III length was consistent, length variability and elongation seem to mainly originate in the iliofemoral trajectory and zone II. The optimal zone I catheter insertion length would be 430 mm. Optimal zone III catheter insertion ranged between 234 and 264 mm. These standardized variable-distance insertion lengths could facilitate safer fluoroscopy-free REBOA in austere, pre-hospital settings.
Authors
- Jan C. van de Voort (ORCID: https://orcid.org/0000-0001-5967-6089)
- Boudewijn L.S. Borger van der Burg (ORCID: https://orcid.org/0000-0002-1199-4042)
- Rigo Hoencamp (ORCID: https://orcid.org/0000-0002-1253-673X)
- Barbara B. Verbeek
Institutions
- Leiden University Medical Center (NL)
- Erasmus MC (NL)
- Ministry of Defence (NL)
- Alrijne Ziekenhuis (NL)
- Erasmus University Rotterdam (NL)
Publication Details
- Journal
- World Journal of Emergency Surgery
- Published
- 2024-08-31
- DOI
- https://doi.org/10.1186/s13017-024-00557-4
- Citations
- 5
- Primary Topic
- Trauma and Emergency Care Studies
- Type
- article
- Field-Weighted Citation Impact
- 3.45
Funders
- Ministry of Defense