The polychromatic retrigger: a new tool to mitigate pileup in spectral photon-counting computed-tomography

High photon rates, like the ones required in clinical computed tomography, can generate pulse pileup in photon-counting detectors, and in extreme cases induce system paralysis. The instant retrigger technology has been developed by DECTRIS Ltd. to counteract such phenomena, effectively making detectors non-paralyzable. The instant retrigger behavior is regulated by a quantity named retrigger time, which originally was the same for all the energy thresholds of the detector. We developed a new ASIC that features an improved version of the instant retrigger (polychromatic retrigger) that allows different retrigger times for each of the four energy thresholds. In this work, we investigate how this feature can be exploited in computed tomography to mitigate pileup artifacts. Using a Monte Carlo method, we modeled a full computed tomography system with four energy thresholds (20, 50, 70 and 90 keV), and we simulated head scans from low (1 <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" id="m1"><mml:mo>×</mml:mo></mml:math> 10 7 cps/mm 2 ) to high (1 <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" id="m2"><mml:mo>×</mml:mo></mml:math> 10 9 cps/mm 2 ) photon count rates for different values of the retrigger time. The quality of the resulting images generally degrades when increasing the photon rates (pileup artifacts) and, at the same time, it becomes strongly dependent on the retrigger time. We show how, even at high incoming rates, it is possible to identify for each energy threshold a retrigger time value that restores almost completely the image quality obtained in the low count-rate scenario. The analysis indicates that the higher the energy threshold, the longer the retrigger time has to be set.

Authors

Publication Details

Journal
Frontiers in Physics
Published
2024-08-30
DOI
https://doi.org/10.3389/fphy.2024.1445538
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The polychromatic retrigger: a new tool to mitigate pileup in spectral photon-counting computed-tomography

Carlo Amato, C. Disch, P. Zambon
Frontiers in Physics
Advanced X-ray and CT Imaging
article

The polychromatic retrigger: a new tool to mitigate pileup in spectral photon-counting computed-tomography

Carlo Amato, C. Disch, P. Zambon
article en

Abstract

High photon rates, like the ones required in clinical computed tomography, can generate pulse pileup in photon-counting detectors, and in extreme cases induce system paralysis. The instant retrigger technology has been developed by DECTRIS Ltd. to counteract such phenomena, effectively making detectors non-paralyzable. The instant retrigger behavior is regulated by a quantity named retrigger time, which originally was the same for all the energy thresholds of the detector. We developed a new ASIC that features an improved version of the instant retrigger (polychromatic retrigger) that allows different retrigger times for each of the four energy thresholds. In this work, we investigate how this feature can be exploited in computed tomography to mitigate pileup artifacts. Using a Monte Carlo method, we modeled a full computed tomography system with four energy thresholds (20, 50, 70 and 90 keV), and we simulated head scans from low (1 <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" id="m1"><mml:mo>×</mml:mo></mml:math> 10 7 cps/mm 2 ) to high (1 <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" id="m2"><mml:mo>×</mml:mo></mml:math> 10 9 cps/mm 2 ) photon count rates for different values of the retrigger time. The quality of the resulting images generally degrades when increasing the photon rates (pileup artifacts) and, at the same time, it becomes strongly dependent on the retrigger time. We show how, even at high incoming rates, it is possible to identify for each energy threshold a retrigger time value that restores almost completely the image quality obtained in the low count-rate scenario. The analysis indicates that the higher the energy threshold, the longer the retrigger time has to be set.

Frontiers in PhysicsVol. 12
Affordable and clean energy
Openalex Percentile: Top 87%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

The polychromatic retrigger: a new tool to mitigate pileup in spectral photon-counting computed-tomography — Carlo Amato, C. Disch, et al. · Frontiers in Physics (2024) | TGRS Research Map | TGRS