Comparing Device-Generated and Calculated Bioimpedance Variables in Community-Dwelling Older Adults

Despite BIA emerging as a clinical tool for assessing older adults, it remains unclear how to calculate whole-body impedance (Z), reactance (Xc), resistance (R), and phase angle (PhA) from segmental values using modern BIA devices that place electrodes on both sides of the body. This investigation aimed to compare both the whole-body and segmental device-generated phase angle (PhA<sub>DG</sub>) with the phase angle calculated using summed Z, Xc, and R from the left, right, and combined sides of the body (PhA<sub>Calc</sub>) and to compare bioelectric variables between sides of the body. A sample of 103 community-dwelling older adults was assessed using a 50 kHz direct segmental multifrequency BIA device. Whole-body PhA<sub>Calc</sub> values were assessed for agreement with PhA<sub>DG</sub> using 2.5th and 97.5th quantile nonparametric limits of agreement and Spearman's rho. Bioelectrical values between sides of the body were compared using Wilcoxon rank and Spearman's rho. A smaller mean difference was observed between PhA<sub>DG</sub> and right PhA<sub>Calc</sub> (-0.004°, <i>p</i> = 0.26) than between PhA<sub>Calc</sub> on the left (0.107°, <i>p</i> = 0.01) and on the combined sides (0.107°, <i>p</i> < 0.001). The sum of Z, R, and PhA<sub>Calc</sub> was significantly different (<i>p</i> < 0.01) between the left (559.66 ± 99.55 Ω, 556.80 ± 99.52 Ω, 5.51 ± 1.5°, respectively) and the right sides (554.60 ± 94.52 Ω, 552.02 ± 94.23 Ω, 5.41 ± 0.8°, respectively). Bilateral BIA values do not appear to be interchangeable when determining whole-body measurements. Present data suggest that using right-sided segmental values would be the most appropriate choice for calculating whole-body bioelectrical variables.

Authors

Institutions

Publication Details

Journal
Sensors
Published
2024-08-30
DOI
https://doi.org/10.3390/s24175626
Citations
5
Primary Topic
Body Composition Measurement Techniques
Type
article
Field-Weighted Citation Impact
1.63

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparing Device-Generated and Calculated Bioimpedance Variables in Community-Dwelling Older Adults

Ladda Thiamwong, Danielle A. Sterner, Jeffrey R. Stout, Joon‐Hyuk Park et al.
5 citations
Sensors
Body Composition Measurement Techniques
1.63
article

Comparing Device-Generated and Calculated Bioimpedance Variables in Community-Dwelling Older Adults

Ladda Thiamwong, Danielle A. Sterner, Jeffrey R. Stout, Joon‐Hyuk Park, Kworweinski Lafontant, David Hideyoshi Fukuda
article en
5 citations

Abstract

Despite BIA emerging as a clinical tool for assessing older adults, it remains unclear how to calculate whole-body impedance (Z), reactance (Xc), resistance (R), and phase angle (PhA) from segmental values using modern BIA devices that place electrodes on both sides of the body. This investigation aimed to compare both the whole-body and segmental device-generated phase angle (PhA<sub>DG</sub>) with the phase angle calculated using summed Z, Xc, and R from the left, right, and combined sides of the body (PhA<sub>Calc</sub>) and to compare bioelectric variables between sides of the body. A sample of 103 community-dwelling older adults was assessed using a 50 kHz direct segmental multifrequency BIA device. Whole-body PhA<sub>Calc</sub> values were assessed for agreement with PhA<sub>DG</sub> using 2.5th and 97.5th quantile nonparametric limits of agreement and Spearman's rho. Bioelectrical values between sides of the body were compared using Wilcoxon rank and Spearman's rho. A smaller mean difference was observed between PhA<sub>DG</sub> and right PhA<sub>Calc</sub> (-0.004°, <i>p</i> = 0.26) than between PhA<sub>Calc</sub> on the left (0.107°, <i>p</i> = 0.01) and on the combined sides (0.107°, <i>p</i> < 0.001). The sum of Z, R, and PhA<sub>Calc</sub> was significantly different (<i>p</i> < 0.01) between the left (559.66 ± 99.55 Ω, 556.80 ± 99.52 Ω, 5.51 ± 1.5°, respectively) and the right sides (554.60 ± 94.52 Ω, 552.02 ± 94.23 Ω, 5.41 ± 0.8°, respectively). Bilateral BIA values do not appear to be interchangeable when determining whole-body measurements. Present data suggest that using right-sided segmental values would be the most appropriate choice for calculating whole-body bioelectrical variables.

SensorsVol. 24(17)
University of Central Florida (US)
National Institute on Aging, National Institute on Minority Health and Health Disparities
Openalex Percentile: Top 16%
Body Composition Measurement Techniques
1.63
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.