Accuracy of Continuous Glucose Monitoring in Hemodialysis Patients With Diabetes
In hemodialysis patients with diabetes, although MARD values were higher than traditional optimal analytic performance thresholds, error grids showed nearly all CGM values were in clinically acceptable zones. Further studies are needed to determine whether CGM improves outcomes in hemodialysis patients.
Authors
- Amy S. You
- Connie M. Rhee (ORCID: https://orcid.org/0000-0002-9703-6469)
- Andrea Daza
- Kamyar Kalantar‐Zadeh (ORCID: https://orcid.org/0000-0002-8666-0725)
- Man Kit Michael Siu
- Danh V. Nguyen (ORCID: https://orcid.org/0000-0002-4025-8239)
- Yoko Narasaki (ORCID: https://orcid.org/0000-0002-2571-3803)
- Renal Amel Peralta
- Alejandra García Novoa (ORCID: https://orcid.org/0000-0002-7141-7570)
Institutions
- University of California, Los Angeles (US)
- University of California, Irvine (US)
- UCLA Medical Center (US)
- Tibor Rubin VA Medical Center (US)
- VA Greater Los Angeles Healthcare System (US)
- Harbor–UCLA Medical Center (US)
Publication Details
- Journal
- Diabetes Care
- Published
- 2024-08-30
- DOI
- https://doi.org/10.2337/dc24-0635
- Citations
- 22
- Primary Topic
- Diabetes Management and Research
- Type
- article
- Field-Weighted Citation Impact
- 7.87
Funders
- Dexcom
- National Institute of Diabetes and Digestive and Kidney Diseases