A cost-effective field-programmable-gate-array-based pulse processor for biomedical imaging applications

This paper presents the design and evaluation of a cost-effective, high-spatial resolution, multichannel, field-programmable gate array (FPGA)-based readout system for Positron Emission Tomography (PET) detectors suitable for clinical and pre-clinical work in cancer and other diseases. A key challenge in PET imaging is the precise measurement of the time of arrival and energy of gamma photons originating from positron annihilation, along with the position of interaction within the detector. These parameters enhance image quality by improving coincidence detection, excluding Compton-scattered photons, and improving system spatial resolution. We developed a 16-channel readout system using single silicon photomultiplier (SiPM) element readout for efficient and precise energy, time, and position measurements. Each channel incorporates an FPGA-based sigma-delta analog-to-digital converter (ΣΔ-ADC) and tapped delay line (TDL) based time-to-digital converter (TDC) for timing and energy estimation. Performance characterization of the system was performed through simulation, electronically generated waveform, and actual PET detector signals from three different single scintillation crystals of face dimensions 3 × 3 mm 2 and depths of 5 and 20 mm coupled to a 3 × 3 mm 2 SiPM. We also constructed and tested a modular PET detector with a 10x10 array of 1.2x1.2x14 mm³ LSO crystals coupled to a 4x4 SiPM array of 3 × 3 mm 2 elements. Best energy resolution was 9.3% at the 511 KeV photopeak and best coincidence timing resolution (CTR) was 210 ± 3.5 ps. Our system balances cost-effectiveness with performance and contributes to developments in PET signal acquisition and detector technology.

Authors

Institutions

Publication Details

Journal
Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment
Published
2024-08-29
DOI
https://doi.org/10.1016/j.nima.2024.169820
Citations
3
Primary Topic
Radiation Detection and Scintillator Technologies
Type
article
Field-Weighted Citation Impact
1.82
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A cost-effective field-programmable-gate-array-based pulse processor for biomedical imaging applications

Felipe Godinez, A. Refaey, Ramsey Derek Badawi
3 citations
Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment
Radiation Detection and Scintillator Technologies
1.82
article

A cost-effective field-programmable-gate-array-based pulse processor for biomedical imaging applications

Felipe Godinez, A. Refaey, Ramsey Derek Badawi
article en
3 citations

Abstract

This paper presents the design and evaluation of a cost-effective, high-spatial resolution, multichannel, field-programmable gate array (FPGA)-based readout system for Positron Emission Tomography (PET) detectors suitable for clinical and pre-clinical work in cancer and other diseases. A key challenge in PET imaging is the precise measurement of the time of arrival and energy of gamma photons originating from positron annihilation, along with the position of interaction within the detector. These parameters enhance image quality by improving coincidence detection, excluding Compton-scattered photons, and improving system spatial resolution. We developed a 16-channel readout system using single silicon photomultiplier (SiPM) element readout for efficient and precise energy, time, and position measurements. Each channel incorporates an FPGA-based sigma-delta analog-to-digital converter (ΣΔ-ADC) and tapped delay line (TDL) based time-to-digital converter (TDC) for timing and energy estimation. Performance characterization of the system was performed through simulation, electronically generated waveform, and actual PET detector signals from three different single scintillation crystals of face dimensions 3 × 3 mm 2 and depths of 5 and 20 mm coupled to a 3 × 3 mm 2 SiPM. We also constructed and tested a modular PET detector with a 10x10 array of 1.2x1.2x14 mm³ LSO crystals coupled to a 4x4 SiPM array of 3 × 3 mm 2 elements. Best energy resolution was 9.3% at the 511 KeV photopeak and best coincidence timing resolution (CTR) was 210 ± 3.5 ps. Our system balances cost-effectiveness with performance and contributes to developments in PET signal acquisition and detector technology.

Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated EquipmentVol. 1069
UC Davis Comprehensive Cancer Center, University of California, Davis (US)
Affordable and clean energy
Openalex Percentile: Top 16%
Radiation Detection and Scintillator Technologies
1.82
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.