Biomechanical effects of endplate sagittal coverage change on cervical disc replacement: a finite element analysis
After the sagittal distance of the posterior edge of the prosthetic endplate was completely covered, the mechanical influence of the entire cervical spine was low. The sagittal distance at the posterior edge of the endplate of different sizes changed the motion pattern and load distribution of the implanted segment to some extent. When the sagittal distance between the prosthesis and the upper endplate was greater than or equal to 3 mm, the mechanical indices of the implanted segment increased significantly, increasing the risk of local tissue injury, especially during extension motion. Compared to the sagittal distance at the posterior edge of the endplate of the lower prosthesis, increasing the sagittal distance at the posterior edge of the endplate of the upper prosthesis has a greater effect on the mechanics of the cervical spine.
Authors
- Lihua Chen (ORCID: https://orcid.org/0000-0003-3369-1875)
- Haiyan Wang (ORCID: https://orcid.org/0000-0001-8728-0719)
- Guangming Xu (ORCID: https://orcid.org/0000-0001-8533-5317)
- Hao Liu (ORCID: https://orcid.org/0000-0002-2992-6232)
Institutions
- Guangzhou University of Chinese Medicine (CN)
- Guangxi University of Chinese Medicine (CN)
- Shenzhen Pingle Orthopedic Hospital (CN)
Publication Details
- Journal
- Frontiers in Bioengineering and Biotechnology
- Published
- 2024-08-29
- DOI
- https://doi.org/10.3389/fbioe.2024.1371548
- Citations
- 2
- Primary Topic
- Cervical and Thoracic Myelopathy
- Type
- article
- Field-Weighted Citation Impact
- 1.15