Alternating magnetic field guiding system for MNP hyperthermia treatment of deep-seated cancers
These results demonstrate proof-of-concept for an endoscopic magnetic field guiding and focusing system capable of delivering clinically relevant AMF from a distance. This innovative approach offers a promising alternative to conventional field delivery methods by directing AMF through the body, concentrating it in the tumor region, reducing eddy currents in surrounding healthy tissue, and avoiding exposure of nearby metallic implants.
Authors
- Ilona Danelyan
- R. Jobava (ORCID: https://orcid.org/0000-0003-3083-0243)
- Robert V. Stigliano
- Fridon Shubitidze (ORCID: https://orcid.org/0000-0001-9674-4810)
- G. Gabriadze (ORCID: https://orcid.org/0000-0002-1210-8014)
- Ian Baker (ORCID: https://orcid.org/0000-0001-6347-6960)
- L. Shoshiashvili
- P. Jack Hoopes
Institutions
- Tbilisi State University (GE)
- Dartmouth College (US)
- Faraday Technology (United States) (US)
Publication Details
- Journal
- International Journal of Hyperthermia
- Published
- 2024-08-29
- DOI
- https://doi.org/10.1080/02656736.2024.2391008
- Citations
- 5
- Primary Topic
- Ultrasound and Hyperthermia Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.92