Exploring the bias: how skin color influences oxygen saturation readings via Monte Carlo simulations
Monte Carlo simulations have demonstrated that different light pathlengths within the tissue contribute to the overestimation of <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"> <mml:mrow><mml:msub><mml:mi>SpO</mml:mi> <mml:mn>2</mml:mn></mml:msub> </mml:mrow> </mml:math> in people with darker skin in reflection mode pulse oximetry. Increasing the SD separation may mitigate the effect of skin tone on <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"> <mml:mrow><mml:msub><mml:mi>SpO</mml:mi> <mml:mn>2</mml:mn></mml:msub> </mml:mrow> </mml:math> readings. These trends were not observed in transmission mode; however, further planned research using more complex models of the tissue is essential.
Authors
- Stephen P. Morgan (ORCID: https://orcid.org/0000-0003-4069-3801)
- Chong Liu (ORCID: https://orcid.org/0000-0002-9521-2983)
- Ricardo Correia (ORCID: https://orcid.org/0000-0001-6405-5340)
- Barrie Hayes‐Gill (ORCID: https://orcid.org/0000-0003-4237-7001)
- Suvvi K. Narayana Swamy
Institutions
- University of Nottingham (GB)
Publication Details
- Journal
- Journal of Biomedical Optics
- Published
- 2024-08-29
- DOI
- https://doi.org/10.1117/1.jbo.29.s3.s33308
- Citations
- 4
- Primary Topic
- Non-Invasive Vital Sign Monitoring
- Type
- article
- Field-Weighted Citation Impact
- 0.74
Funders
- Medical Research Council
- Engineering and Physical Sciences Research Council