Diagnostic Accuracy of the Persyst Automated Seizure Detector in the Neonatal Population
The Persyst neonatal SDA demonstrates high diagnostic accuracy in identifying neonatal seizures; comparable to the accuracy of the standard Persyst SDA in adult populations, other neonatal SDAs, and amplitude integrated EEG (aEEG). Overdiagnosis of seizures is a risk, particularly from cEEG recording artefact. To fully examine its clinical utility, further investigation of the Persyst neonatal SDA's accuracy is required, as well as confirming the optimal seizure probability thresholds in a larger patient cohort.
Authors
- Stewart Boyd (ORCID: https://orcid.org/0000-0001-7363-2453)
- Eleanor Duckworth
- Maria Chalia (ORCID: https://orcid.org/0000-0003-2441-6564)
- Daniyal Motan
- Ronit Pressler (ORCID: https://orcid.org/0000-0002-2905-6839)
- Kitty Howse
Institutions
- Great Ormond Street Hospital (GB)
- University College London (GB)
Publication Details
- Journal
- Journal of Integrative Neuroscience
- Published
- 2024-08-19
- DOI
- https://doi.org/10.31083/j.jin2308150
- Citations
- 2
- Primary Topic
- Neonatal and fetal brain pathology
- Type
- article
- Field-Weighted Citation Impact
- 2.22