Development and Integration of an Ultraminiaturized Gas Chromatograph Prototype Based on Lab-on-a-Chip Microelectromechanical Systems for Space Exploration Missions

We developed, integrated, and tested an ultraminiaturized gas chromatograph prototype for space applications based on microelectromechanical system (MEMS) preconcentrator, column, and detector subunits. The fluidic interface, the mechanical design, and the electronic system were all designed to be easily transposable to space instrumentation. The response linearity, reproducibility as well as the preconcentrator, column, and detector performances were demonstrated using a mixture of n-pentane, n-hexane, benzene, and toluene. Using the column, the response linearity of the nano-gravimetric detector (NGD) was observed over 1 to 2 orders of magnitude with a high R2 value of 0.99. The retention time reproducibility was evaluated with a variation lower than 0.2%. The column separation performances were also qualified in terms of number of theoretical plates (N) and height equivalent to a theoretical plate (HETP) with 8275 ± 45 plates (equivalent to 1655 ± 9 plates per meter) and 0.6 mm on average, respectively. The preconcentrator performance for trapping and desorption was also validated. The NGD showed high sensitivity, and the limit of detection was evaluated to be about 3.1 pmol for toluene, which is very suitable for space exploration. The analytical performance of the first μGC-MEMS prototype was already satisfying and adapted for space studies. Nevertheless, some ways of optimization were identified (like back-flush injection for the preconcentrator or the use of valves more suited to space applications) to improve both performances and instrument robustness, and they will be tested in the future with a second version of our prototype.

Authors

Institutions

Publication Details

Journal
ACS Earth and Space Chemistry
Published
2024-08-29
DOI
https://doi.org/10.1021/acsearthspacechem.4c00111
Citations
4
Primary Topic
Planetary Science and Exploration
Type
article
Field-Weighted Citation Impact
1.33

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development and Integration of an Ultraminiaturized Gas Chromatograph Prototype Based on Lab-on-a-Chip Microelectromechanical Systems for Space Exploration Missions

Valérie Peulon‐Agasse, Vincent Guerrini, Guillaume Rioland, Arnaud Philippart et al.
4 citations
ACS Earth and Space Chemistry
Planetary Science and Exploration
1.33
article

Development and Integration of an Ultraminiaturized Gas Chromatograph Prototype Based on Lab-on-a-Chip Microelectromechanical Systems for Space Exploration Missions

Valérie Peulon‐Agasse, Vincent Guerrini, Guillaume Rioland, Arnaud Philippart, Frédéric Ferreira, Pascal Cardinaël, Cyril Szopa, Fabrice Bertrand, Malak Rizk-Bigourd, A. Buch, Audrey Boco, Jean-Pierre Pineau, David Coscia
article en
4 citations

Abstract

We developed, integrated, and tested an ultraminiaturized gas chromatograph prototype for space applications based on microelectromechanical system (MEMS) preconcentrator, column, and detector subunits. The fluidic interface, the mechanical design, and the electronic system were all designed to be easily transposable to space instrumentation. The response linearity, reproducibility as well as the preconcentrator, column, and detector performances were demonstrated using a mixture of n-pentane, n-hexane, benzene, and toluene. Using the column, the response linearity of the nano-gravimetric detector (NGD) was observed over 1 to 2 orders of magnitude with a high R2 value of 0.99. The retention time reproducibility was evaluated with a variation lower than 0.2%. The column separation performances were also qualified in terms of number of theoretical plates (N) and height equivalent to a theoretical plate (HETP) with 8275 ± 45 plates (equivalent to 1655 ± 9 plates per meter) and 0.6 mm on average, respectively. The preconcentrator performance for trapping and desorption was also validated. The NGD showed high sensitivity, and the limit of detection was evaluated to be about 3.1 pmol for toluene, which is very suitable for space exploration. The analytical performance of the first μGC-MEMS prototype was already satisfying and adapted for space studies. Nevertheless, some ways of optimization were identified (like back-flush injection for the preconcentrator or the use of valves more suited to space applications) to improve both performances and instrument robustness, and they will be tested in the future with a second version of our prototype.

ACS Earth and Space ChemistryVol. 8(9)
Centre National de la Recherche Scientifique (FR), Université de Versailles Saint-Quentin-en-Yvelines (FR), Université Paris-Saclay (FR), Centre National d'Études Spatiales (FR), Sorbonne Université (FR), Normandie Université (FR), CentraleSupélec (FR), Institut Pierre-Simon Laplace (FR), Laboratoire atmosphères, milieux, observations spatiales (FR)
Centre National d’Etudes Spatiales
Openalex Percentile: Top 20%
Planetary Science and Exploration
1.33
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.