Increasing the Beam Width and Intensity with Refraction Power Effect Using a Combination of Beam Mirrors and Concave Mirrors for Surgical-Fluorescence-Emission-Guided Cancer Monitoring Method

The primary goal during cancer removal surgery is to completely excise the malignant tumor. Because the color of the tumor and surrounding tissues is very similar, it is difficult to observe with the naked eye, posing a risk of damaging surrounding blood vessels during the tumor removal process. Therefore, fluorescence emission is induced using a fluorescent contrast agent, and color classification is monitored through camera imaging. LEDs must be irradiated to generate the fluorescent emission electromotive force. However, the power and beam width of the LED are insufficient to generate this force effectively, so the beam width and intensity must be increased to irradiate the entire lesion. Additionally, there should be no shaded areas in the beam irradiation range. This paper proposes a method to enhance the beam width and intensity while eliminating shadow areas. A total reflection beam mirror was used to increase beam width and intensity. However, when the beam width increased, a shadow area appeared at the edge, limiting irradiation of the entire lesion. To compensate for this shadow area, a concave lens was combined with the beam mirror, resulting in an increase in beam width and intensity by more than 1.42 times and 18.6 times, respectively. Consequently, the beam width reached 111.8°, and the beam power was 13.6 mW. The proposed method is expected to be useful for observing tumors through the induction of fluorescence emission during cancer removal surgery or for pathological examination in the pathology department.

Authors

Institutions

Publication Details

Journal
Sensors
Published
2024-08-25
DOI
https://doi.org/10.3390/s24175503
Primary Topic
Photodynamic Therapy Research Studies
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Increasing the Beam Width and Intensity with Refraction Power Effect Using a Combination of Beam Mirrors and Concave Mirrors for Surgical-Fluorescence-Emission-Guided Cancer Monitoring Method

Ji-Na Park, Ki‐Cheol Yoon, Minchan Kim, Kwang Gi Kim et al.
Sensors
Photodynamic Therapy Research Studies
article

Increasing the Beam Width and Intensity with Refraction Power Effect Using a Combination of Beam Mirrors and Concave Mirrors for Surgical-Fluorescence-Emission-Guided Cancer Monitoring Method

Ji-Na Park, Ki‐Cheol Yoon, Minchan Kim, Kwang Gi Kim, Jeong-Min Seo, Sangyun Lee, Seung Yeob Ryu
article en

Abstract

The primary goal during cancer removal surgery is to completely excise the malignant tumor. Because the color of the tumor and surrounding tissues is very similar, it is difficult to observe with the naked eye, posing a risk of damaging surrounding blood vessels during the tumor removal process. Therefore, fluorescence emission is induced using a fluorescent contrast agent, and color classification is monitored through camera imaging. LEDs must be irradiated to generate the fluorescent emission electromotive force. However, the power and beam width of the LED are insufficient to generate this force effectively, so the beam width and intensity must be increased to irradiate the entire lesion. Additionally, there should be no shaded areas in the beam irradiation range. This paper proposes a method to enhance the beam width and intensity while eliminating shadow areas. A total reflection beam mirror was used to increase beam width and intensity. However, when the beam width increased, a shadow area appeared at the edge, limiting irradiation of the entire lesion. To compensate for this shadow area, a concave lens was combined with the beam mirror, resulting in an increase in beam width and intensity by more than 1.42 times and 18.6 times, respectively. Consequently, the beam width reached 111.8°, and the beam power was 13.6 mW. The proposed method is expected to be useful for observing tumors through the induction of fluorescence emission during cancer removal surgery or for pathological examination in the pathology department.

SensorsVol. 24(17)
Gachon University (KR), Dongnam Health University (KR), Gachon University Gil Medical Center (KR)
Korea Medical Device Development Fund, Ministry of Trade, Industry and Energy, Ministry of Food and Drug Safety, Ministry of Science and ICT, South Korea
Good health and well-being
Openalex Percentile: Top 75%
Photodynamic Therapy Research Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.